Step 2, About the hr/pr control menu about the spo, Control menu – Welch Allyn 901061 Propaq LT Vital Signs Monitor - User Manual User Manual
Page 38
34
Overview of monitor operation
Welch Allyn Propaq LT Vital Signs Monitor
2.
Scroll (using
) to HR/PR Tone, and press either
or
as many times as needed to
highlight Off.
3.
Press
or
to exit the control screen and return to the vital-signs display.
About the HR/PR control menu
About the SpO
2
control menu
Note
When you change a setting (for example, by turning off an alarm limit or by
increasing or decreasing an alarm limit), the change takes effect immediately.
Note
When you exit a control menu, the values displayed at the time you exit are the
values in effect for the monitor. If you change a parameter setting and then decide
before exiting the control menu to keep the previous setting values, you must
return the parameters to the original values before you exit the control menu.
Parameter
Options
Parameter
Options
Upper Alarm
Off On
Lower Alarm
Off On
Upper Limits
Lower Limits
Adult
Pediatric
Neonate
27 - 300 beats/minute
27 - 300 beats/minute
27 - 300 beats/minute
Adult
Pediatric
Neonate
25 - 298 beats/minute
25 - 298 beats/minute
25 - 298 beats/minute
HR/PR Tone
Off Low Med High
Selected Source
ECG SpO
2
Parameter
Options
Parameter
Options
SpO
2
Monitoring
Off On Standby
HR/PR Tone
Off Low Med High
Upper Alarm
Off On
Lower Alarm
Off On
Upper Limit
Lower Limit
Adult
Pediatric
Neonate
52% - 100%
52% - 100%
52% - 100%
Adult
Pediatric
Neonate
50% - 98%
50% - 98%
50% - 98%