62

Standalone monitoring

Welch Allyn Propaq LT Vital Signs Monitor

-

the patient is in cardiac arrest or shock

-

the patient has hypotension, severe vasoconstriction, severe anemia, or
hypothermia

If poor perfusion affects performance for an adult, consider using the adult nasal
sensor.

If a sensor is connected and the sensor light does not come on within 3 seconds:

-

Verify that SpO

2

is turned on (

Figure 67

on page 62)

-

Replace the sensor

If excessive or prolonged patient movement interferes with measurements,
consider the following possible solutions:

-

be sure the sensor is secure and properly applied

-

use a new sensor with fresh adhesive backing

-

select a different type of sensor

-

move the sensor to a less active site

6.

Periodically verify that the sensor remains properly positioned on the patient.

About SpO

2

spot check

When SpO

2

is turned on, the monitor generates an alarm condition whenever SpO

2

readings are interrupted, such as when the sensor is disconnected from the patient after
the monitor begins taking SpO

2

readings.

Using the SpO

2

Spot Check feature, however, you can take any number of spot SpO

2

readings at random intervals, attaching and detaching the sensor repeatedly without
generating alarms.

To prepare to take a spot check reading

1.

Highlight SpO2 and press

.

2.

Highlight SpO2 Monitoring, press

to highlight Off, and press

to return to the

main screen.

Figure 67. Turning off SpO

2

monitoring

‘SpO

2

’ has changed to ‘SpO

2

@ (time)’. Spot checks are now enabled.

Note

The Spot Check feature is available only if it is enabled in the monitor
configuration. Refer to

“Using the Configuration Utility to configure the monitor”

on page 116.

SpO2 Monitoring

SpO2 Monitoring
Upper Alarm

Upper Alarm
Upper Limit

Upper Limit
Lower Alarm

Lower Alarm
Lower Limit

Lower Limit
HP/PR Tone

HP/PR Tone

SpO2

SpO2

Off

Off
Off

Off

Off

Off

Off

Off

On

On
On

On

On

On

Low

Low

Med

Med High

High

100

100

90

90

Standby

Standby

Press

to move the

highlight to Off.