Standard laryngoscope blade assemblies
Directions for use

Intended use

The rigid laryngoscope is a device used to examine and visualize a patient’s airway and aid
placement of a tracheal tube.

About this document

This directions for use applies to Welch Allyn reusable standard laryngoscope blade
assemblies MacIntosh REF: 6904X, English MacIntosh REF: 6924X, and Miller REF:
6804X and REF 68470. Welch Allyn reusable standard laryngoscope blade assemblies may be
used with Welch Allyn standard laryngoscope handles 60200, 60300, 60400, 60305, and 60710.

Warnings and cautions

WARNING: Reusable fiber optic laryngoscope blade assemblies must be reprocessed after
each use.

WARNING: The reprocessing procedure and the equipment and materials described must
be followed and conducted by persons trained and familiar with medical device
reprocessing.

WARNING: Consult cleaning and disinfecting agent manufacturer instructions for their
proper preparation and use.

WARNING: Repeated reprocessing may degrade elements of the device. Follow inspection
procedures to assure damage has not occurred to the blade assembly.

WARNING: Only trained personnel shall use a laryngoscope for intubation.

WARNING: Do not reprocess laryngoscope blade assemblies used on patients with
Creutzfeldt-Jakob disease. These instruments should be discarded.

WARNING: Discard any component that shows evidence of damage or deterioration.

WARNING: Do not modify this equipment. Any modification of this equipment can lead to
patient injury. Any modification of this equipment voids the product warranty.

WARNING: Personnel shall follow their facility policies and procedures and wear appropriate
personal protective equipment when handling potentially contaminated equipment.

WARNING: Laryngoscope equipment is not suitable for use in intense magnetic fields.

WARNING: Lamps if left illuminated, could generate sufficient heat to cause burns.

CAUTION: Only use lamps specified. Failure to follow these instructions may cause damage
or poor performance of this device.

CAUTION: Only use neutral pH (6.0-8.0) enzymatic type cleaners.

CAUTION: Only use neutral pH (6.0-8.0) ortho-phthalahyde (OPA) or glutaraldehyde
disinfectants.

CAUTION: Do not use other cleaners or disinfectants.

CAUTION: Do not remove lamp during reprocessing. Damage to device could occur.

CAUTION: REF 68470 is not compatible with any sterilization process.