Welch Allyn 901061 Propaq LT Vital Signs Monitor - User Manual User Manual
Page 29

Directions for Use
Overview of monitor operation
25
Leads I, II, and III are viewable. The waveform for the lead selected at the monitor is
displayed on the large color display. Lead V cannot be displayed.
Figure 21. 9-waveform view, 3-lead cable
Leads I, II, and III are viewable. The waveform for the lead selected at the monitor is
displayed on the large color display. Leads V, aVR, aVL, and aVF cannot be displayed.
Installing the interface box
The Large Color Display Interface receives DC power from the AC power adapter and
vital-signs data from the cradle. It provides DC power to the cradle and the video signal to
the large display.
Figure 22. Interface box connections
When the interface box is attached to the cradle, the default power and USB cables
connect the interface box and the cradle. (See
If the interface box and the cradle are not attached, longer cables are used.
The AC power adapter is shipped with the cradle.
Note
When using a 5-lead cable with all electrodes attached, only leads II and V are
available in the 4-waveform view.
Note
When using a 5-lead cable with all electrodes attached, leads I, II, III, V, aVR, aVL,
and aVF are available in the 9-waveform view.
Note
The AC power adapter used with the Propaq CS monitor or the Propaq Encore
monitor will not power the interface box.
DC power out
(to cradle)
VGA out to
(large color display)
DC power in
(from AC power adapter)
Fuse
USB in
(from cradle)
DC power indicator