Directions for Use

Overview of monitor operation

25

Leads I, II, and III are viewable. The waveform for the lead selected at the monitor is
displayed on the large color display. Lead V cannot be displayed.

Figure 21. 9-waveform view, 3-lead cable

Leads I, II, and III are viewable. The waveform for the lead selected at the monitor is
displayed on the large color display. Leads V, aVR, aVL, and aVF cannot be displayed.

Installing the interface box

The Large Color Display Interface receives DC power from the AC power adapter and
vital-signs data from the cradle. It provides DC power to the cradle and the video signal to
the large display.

Figure 22. Interface box connections

When the interface box is attached to the cradle, the default power and USB cables
connect the interface box and the cradle. (See

Figure 17

on page 23.)

If the interface box and the cradle are not attached, longer cables are used.

The AC power adapter is shipped with the cradle.

Note

When using a 5-lead cable with all electrodes attached, only leads II and V are
available in the 4-waveform view.

Note

When using a 5-lead cable with all electrodes attached, leads I, II, III, V, aVR, aVL,
and aVF are available in the 9-waveform view.

Note

The AC power adapter used with the Propaq CS monitor or the Propaq Encore
monitor will not power the interface box.

DC power out
(to cradle)

VGA out to
(large color display)

DC power in
(from AC power adapter)

Fuse

USB in
(from cradle)

DC power indicator