beautypg.com

Overview of monitor operation, Turning on the monitor, 2 - overview of monitor operation – Welch Allyn 901061 Propaq LT Vital Signs Monitor - User Manual User Manual

Page 17

background image

2

13

Overview of monitor operation

Turning on the monitor

The monitor runs through an operational self-test each time it is powered on. Always
verify that it follows the power-on sequence shown below. When

Step 3

completes as

described, the monitor has verified that the visual and audible alarm indicators are
working properly. If

Step 4

does not complete as described, remove the monitor from

service and have it examined by a qualified service person.

To turn on the monitor

1.

Press and hold

until a tone sounds.

2.

A tone sounds as the green, yellow, and red lights turn on and off. This occurs twice.

3.

The splash screen appears, identifying the product and displaying the message
“Diagnostics in progress”. The green light turns on and off quickly while the low tone
sounds, the yellow light turns on and off quickly while the medium tone sounds, and
the red light turns on and off quickly while the loud tone sounds.

Figure 6. Splash screen

4.

The splash screen is replaced by one of the power-on screens shown here:

Figure 7. Power-up screens

To start monitoring from this point, see

“Standalone monitoring”

on page 47. To practice

using the monitor, see

“Using demo mode”

on page 40.

Application Version 1.00.00

Application Version 1.00.00

Bootcore Version 1.00.00

Bootcore Version 1.00.00

Propaq

Propaq

®

LT

LT

Diagnostics in progress

Diagnostics in progress

Welch Allyn 2005

Welch Allyn 2005

c

Select ( ) to Start a New Patient.

Select ( ) to Start a New Patient.

No data saved.

No data saved.

Start New Patient

Start New Patient

Info

Info

Demo

Demo

Portland Westside Hospital

Portland Westside Hospital

Emergency Department

Emergency Department

Select ( ) to delete data

Select ( ) to delete data

and start a new patient.

and start a new patient.

Portland Westside Hospital

Portland Westside Hospital

Emergency Department

Emergency Department

Note!

Note!

There is patient data stored for

There is patient data stored for

Hall, Robert E. ID: 3456187

Hall, Robert E. ID: 3456187

Start New Patient

Start New Patient

Continue Patient

Continue Patient

Info

Info Demo

Demo