beautypg.com

Default limits – Welch Allyn 901061 Propaq LT Vital Signs Monitor - User Manual User Manual

Page 154

background image

150

Specifications

Welch Allyn Propaq LT Vital Signs Monitor

Default limits

Patient alarm
high

-

priority limit violation

[100 ms on, 80 ms off, 100 ms on, 80 ms off, 100 ms on, 260 ms off, 100 ms
on, 80 ms off, 100 ms on, 1 sec off, 100 ms on, 80 ms off, 100 ms on, 80 ms
off, 100 ms on, 260 ms off, 100 ms on, 80 ms off, 100 ms on, 4 sec off];
repeat.

Equipment alert
medium priority

[160 ms on, 200 ms off, 160 ms on, 200 ms off, 160 ms on, 3.96 sec off];
repeat.

Limits

Settable on all parameters
Separate adult/pediatric/neonate mode settable limits

Alarm control

Automatic preset or manual settings

Alarm priority

High
Medium
Low

Patient alarms (Life-threatening)
Equipment alerts
Alarm off

Alarm on tachycardias

Most tachycardias alarm in less than 8 seconds. These include AAMI 3.1.2.1
part f. waveforms. Certain multifocal tachycardias can initially alarm as “low
rate.”

Alarm hold-off period

HR/PR
% SpO

2

RR

3 seconds
10 seconds
5 seconds

To keep false alarms to a minimum, the monitor briefly delays or “holds off”
triggering audible and visual alarms for limit violations for these vital signs.
After the hold-off period begins, if the monitor detects that the vital sign has
returned to acceptable limits, the monitor cancels the alarm. The next time a
limit is violated, the monitor starts a new hold-off.

Audio alarm hold-off with Acuity

When a monitor is connected to Acuity, the audio alarms at the monitor can
be delayed up to 4 minutes and 15 seconds. The delay time is selected in
Acuity software at the time of Acuity installation. Visual alarm indications
are not delayed.
Acuity has a default audio alarm hold-off of 11 seconds.

Alarms suspend

When ‘alarms suspend’ is enabled, the monitor alarm tones for all
parameters can be suspended on the floor for a configurable period: 90
seconds, 2 minutes, 3 minutes, 4 minutes (default), 5 minutes, 10 minutes, 15
minutes, or 60 minutes.

Patient out

-

of

-

range;

transmitter failure

When a monitor is used with Acuity, an equipment alert is generated
whenever the monitor fails to communicate with Acuity after a connection
has been successfully established. In addition, the ‘No Acuity’ icon is
displayed on the monitor.

Battery failure

An equipment alert is generated before the monitor battery becomes
exhausted.

Snapshot/event mark

When

is pressed, the monitor records a Snapshot. If the monitor is

connected to Acuity, the monitor also sends a message to Acuity so Acuity
can mark and print it.

Characteristic

Specification

Upper

Lower

HR/PR

Adult

120 beats/min

Pediatric

150 beats/min

Neonate

200 beats/min

Adult

50 beats/min

Pediatric

50 beats/min

Neonate

100 beats/min

Resp

Adult

30 breaths/min

Pediatric

45 breaths/min

Neonate

60 breaths/min

Adult

5 breaths/min

Pediatric

10 breaths/min

Neonate

10 breaths/min