Directions for Use

Compliance

163

Monitor, battery-operated

The EMC specifications listed on pages

163

to

174

apply to Propaq LT monitors using any

ECG cable listed in the accessory list or at

www.welchallyn.com

. For other EMC

specifications, see

“Monitor, battery-operated, with cable 008-0799-00/01”

on page 175.

Guidance and manufacturer’s declaration—electromagnetic emissions

The battery-operated Propaq LT Series monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of the monitor should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment—guidance

RF emissions
CISPR 11

Group 1

The battery-operated Propaq LT Series monitor uses RF
energy only for its internal function.

a

Therefore, its RF

emissions are very low and are not likely to cause any
interference in nearby electronic equipment.

a.

The battery-operated Propaq LT Series monitor contains either a 5-GHz orthogonal frequency-division multiplexing transmitter or a
2.4

-

GHz frequency

-

hopping spread

-

spectrum transmitter for the purpose of wireless communication. The radio is operated according

to the requirements of various agencies, including FCC 47 CFR 15.247 and R&TTE Directive (1995/5/EC). The radio is excluded from
the EMC requirements of 60601

-

1

-

2:2001, but should be considered when addressing possible interference issues between this and

other devices.

RF emissions
CISPR 11

Class B

The battery-operated Propaq LT Series monitor is suitable for
use in all establishments, including domestic establishments
and those directly connected to the public low

-

voltage power

supply network that supplies buildings used for domestic
purposes.

Harmonic emissions
IEC 61000

-

3

-

2

No connection to mains

(battery

-

operated)

Voltage fluctuations/flicker
emissions
IEC 61000

-

3

-

3

No connection to mains

(battery

-

operated)