Specifications
145
Pacer pulse rejection
Does not count as heartbeats approximately 95% of pacemaker pulses within pacer
pulse detection range*, with or without AAMI (EC 13
-
2002) tails of 4, 8, 15, 25, 50, 75,
or 100 ms decay time constant, whose tail amplitudes meet either EC 13’s method A or
method B criteria for single pacer pulses, or A
-
V sequential pulses (150 ms and 250 ms
separation), all per AAMI tests 4.1.4.1 and 4.1.4.2.
* Rejects method B overshoots for [pulse width] X [pulse amplitude] < 3 X 10
-4
V-sec.
Pacer pulse detector minimum
slew rate detection threshold
2.5 V/sec RTI, per EC13:2002, 4.1.4.3
ECG trace duration
3 seconds in single or double trace display
6 seconds in cascade
Response to Irregular Rhythm (AAMI specification EC13
-
2002, 4.1.2.1. Part e.)
Ventricular bigeminy (VB)
78 to 82 beats/min (80 beats/min expected)
Slow alternating VB
57 to 65 beats/min (60 beats/min expected)
Rapid alternating VB
118 to 123 beats/min (120 beats/min expected)
Bidirectional systole
88 to 93 beats/min (90 beats/min expected)
1 mV ventricular tachycardia
197 to 198 beats/min (206 beats/min expected)
2 mV ventricular tachycardia
193 to 197 beats/min (195 beats/min expected)
- GS 777 Wall Transformer - User Manual (1 page)
- 7114x Desk Charger - User Manual (8 pages)
- Mounting Bracket Replacement Kit for 7670-12 Mobile Stand with Mounting for 767XX and 777XX - User Manual (2 pages)
- 767 Diagnostic System - User Manual (16 pages)
- 767 Diagnostic System - User Manual (136 pages)
- 118 Series PanOptic Ophthalmoscope - User Manual (28 pages)
- 118 Series PanOptic Ophthalmoscope - User Manual (26 pages)
- VS100 Welch Allyn Vision Screener - User Manual (34 pages)
- SureSight Vision Screener - User Manual (36 pages)
- SureSight Vision Screener - User Manual (32 pages)
- Connex Integrated Wall System - User Manual (161 pages)
- SureSight Autorefractor - User Manual (36 pages)
- 12500 Binocular Indirect Ophthalmoscope Power Source - User Manual (12 pages)
- TM286 Auto Tymp - User Manual (72 pages)
- TM 262 Auto Tymp - User Manual (92 pages)
- MicroTymp 3 portable tympanometric instrument - User Manual (76 pages)
- Audioscope 3 Portable Screening Audiometer - User Manual (32 pages)
- AM282 Audiometer - User Manual (32 pages)
- AM 232 Manual Audiometer - User Manual (38 pages)
- Digital MacroView Otoscope - User Manual (32 pages)
- Digital MacroView Otoscope - User Manual (476 pages)
- OAE Hearing Screener - User Manual (62 pages)
- OAE Hearing Screener - User Manual (56 pages)
- OAE Data Manager - User Manual (39 pages)
- Ear Wash System 29350 - User Manual (28 pages)
- Standard laryngoscope blade assemblies - User Manual (6 pages)
- Standard laryngoscope handles - User Manual (6 pages)
- Rechargeable laryngoscope handles - User Manual (8 pages)
- Fiber optic laryngoscope handles - User Manual (7 pages)
- Fiber optic laryngoscope blade assemblies - User Manual (6 pages)
- Original Harvey and Harvey DLX Double and Triple Head Stethoscopes - User Manual (28 pages)
- Harvey Elite Stethoscope - User Manual (2 pages)
- Professional Stethoscope - User Manual (2 pages)
- EXPENDABLE ILLUMINATOR - User Manual (2 pages)
- KleenSpec Single Use Vaginal Speculum - User Manual (2 pages)
- KleenSpec Vaginal Specula Illumination System - User Manual (20 pages)
- KleenSpec 790 Series Cordless Illumination System - User Manual (32 pages)
- KleenSpec 790 Series Cordless Illumination System - User Manual (222 pages)
- Video Colposcope - User Manual (48 pages)
- Video Colposcope - User Manual (400 pages)
- Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, Accessories - Cleaning, Disinfection, and Sterilization - User Manual (12 pages)
- 6V Power Supply, Rectal Light Handle - User Manual (240 pages)
- Fl-100 Intubating Fiberscope - User Manual (32 pages)
- EpiScope Skin Surface Microscope - User Manual (2 pages)
- 719 Series Lithium Ion Handle - User Manual (2 pages)