Welch Allyn 901061 Propaq LT Vital Signs Monitor - User Manual User Manual
Page 114

110
Monitor configuration
Welch Allyn Propaq LT Vital Signs Monitor
NIBP pediatric
Auto Interval (minutes)
15
1, 2, 3, 5, 10, 15, 30, 60
Monitor, Configuration
Utility
Mode
Manual
Manual, Auto
Monitor, Acuity
Smartcuf
Off
On, Off
Monitor, Acuity,
Configuration Utility
Systolic Upper Alarm Limit
145 mmHg
19.3 kPa
32 - 160 mmHg
4.3 - 21.3 kPa
Monitor, Acuity
Systolic Lower Alarm Limit
75 mmHg
10.0 kPa
30 - 158 mmHg
4.0 - 21.1 kPa
Monitor, Acuity
Diastolic Upper Alarm Limit
100 mmHg
13.3 kPa
17 - 130 mmHg
2.3 - 17.3 kPa
Monitor, Acuity
Diastolic Lower Alarm Limit
35 mmHg
4.7 kPa
15 - 128 mmHg
2.0 - 17.1 kPa
Monitor, Acuity
Mean Upper Alarm Limit
110 mmHg
14.7 kPa
17 - 140 mmHg
2.3 - 18.7 kPa
Monitor, Acuity
Mean Lower Alarm Limit
50 mmHg
6.7 kPa
15 - 138 mmHg
2.0 - 18.4 kPa
Monitor, Acuity
NIBP neonatal
Auto Mode Interval (minutes)
15
1, 2, 3, 5, 10, 15, 30, 60
Monitor, Configuration
Utility
Mode
Manual
Manual, Auto
Monitor, Acuity
Smartcuf
Off
On, Off
Monitor, Acuity,
Configuration Utility
Systolic Upper Alarm Limit
100 mmHg
13.3 kPa
27 - 120 mmHg
3.6 - 16.0 kPa
Monitor, Acuity
Systolic Lower Alarm Limit
50 mmHg
6.7 kPa
25 - 118 mmHg
3.33 - 15.7 kPa
Monitor, Acuity
Diastolic Upper Alarm Limit
70 mmHg
9.3 kPa
12 - 105 mmHg
1.6 - 14.0 kPa
Monitor, Acuity
Diastolic Lower Alarm Limit
30 mmHg
4.0 kPa
10 - 103 mmHg
1.3 - 13.7 kPa
Monitor, Acuity
Mean Upper Alarm Limit
80 mmHg
10.7 kPa
12 - 110 mmHg
1.6 - 14.7 kPa
Monitor, Acuity
Mean Lower Alarm Limit
35 mmHg
4.7 kPa
10 - 108 mmHg
1.3 - 14.4 kPa
Monitor, Acuity
SpO
2
general
Enable Spot Checks
Yes
Yes, No
Configuration Utility
ParamSet for Upper SpO
2
5%
5% - 10%
Configuration Utility
ParamSet for Lower SpO
2
5%
5% - 10%
Configuration Utility
SpO
2
adult
Upper Alarm Limit
100%
52% - 100%
Monitor, Acuity
Lower Alarm Limit
90%
50% - 98%
Monitor, Acuity
Table 10. Parameter configuration matrix (continued)
Parameter
Factory setting
Possible values
Where set