Specifications, A - specifications – Welch Allyn 901061 Propaq LT Vital Signs Monitor - User Manual User Manual
Page 147

A
143
Specifications
Flat-panel display for the Large Color Display Interface
. . . . . . . . . . . . . . . . . . . . . . . . . 153
Environmental
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154
ECG
Characteristic
Specification
Regulatory standards
Meets ANSI/AAMI EC13-2002 and EN 60601-1:1990 + A1 + A2
Connector
Hypertronics 9
-
pin female D01 latching connector
Selectable leads
3-lead cable
5-lead cable
Displayable
I, II or III
Internally provides
I, II or III
Displayable
I, II, III, aV
R
, aV
L
, aV
F,
V
Internally provides
I/II, III and V
Cable type detection
Automatic detection and differentiation of 3
-
or 5-lead cable
Lead fault indicator
Displayed chest icon with flashing indicator for each electrode
ECG Size (sensitivity)
0.2, 0.5, 1, 2, 4, and 8 mV/cm
Display sweep speed
25 mm/sec
Bandwidth (monitor display)
0.5 to 80 Hz
Monitor Mode
0.05 to 80 Hz
Extended Mode
Bandwidth (to Acuity)
0.05 to 80 Hz
Monitor or Extended Mode
Notch filters
50 and 100 Hz, 60 and 120 Hz, or OFF, selectable at Acuity or at the monitor
Sample rate
363.64 Hz (181.82 Hz to Acuity)
Input protection
.
Electrosurgery interference
suppression
Included on all vectors when Welch Allyn-approved cables are used.