beautypg.com

Menus, Figure 29, Using control menus – Welch Allyn 901061 Propaq LT Vital Signs Monitor - User Manual User Manual

Page 37: Example: using a control menu

background image

Directions for Use

Overview of monitor operation

33

Menus

Using control menus

Figure 29. SpO

2

control menu (example)

A control menu includes a topic name for the current context (for example, SpO2); a
column of parameters with one highlighted (for example, SpO2 Monitoring); and a
column of options, with one item in each set of options highlighted (for example, On, On,
100, On, 90, Low).

The blue highlight indicates the parameter currently enabled for modification.

The green highlights indicate the current settings for all parameters in the menu.

At the bottom of the screen for all control menus are links to Exit, Trends, Snapshots,
and Setup.

Example: Using a control menu

Using the example (

Figure 29

), you would do the following to raise the SpO

2

lower alarm

limit to 95 (

Step 1

) and shut off the HR/PR tone (

Step 2

):

1.

With SpO2 Monitoring highlighted, scroll (using

) to highlight Lower Limit, and

press

as many times as needed to raise this alarm limit to 95.

Exit

Return to the vital-signs display.

Trends

View a tabular history.

Snapshots

View a series of 21-second waveform snapshots of the current patient’s
vital signs.

Setup

Access the setup menu. (See

“To access the setup menus”

on page 37.)

%

HALL, ROBERT E.

HALL, ROBERT E.

3456187

3456187

3:00:06P

3:00:06P

Adult

Adult

Rm 239

Rm 239

II 1mV/cm

II 1mV/cm

Exit

Exit

Trends

Trends

Snapshots

Snapshots

Setup

Setup

SpO2 Monitoring

SpO2 Monitoring
Upper Alarm

Upper Alarm
Upper Limit

Upper Limit
Lower Alarm

Lower Alarm
Lower Limit

Lower Limit
HP/PR Tone

HP/PR Tone

SpO2

SpO2

Off

Off
Off

Off

Off

Off

Off

Off

On

On

On

On

On

On

Low

Low

Med

Med High

High

100

100

90

90

Standby

Standby

80

80

HR/min

HR/min

140/78

140/78

NIBP mmHg (102)

NIBP mmHg (102)

Resp/min

Resp/min

SpO2

SpO2

12

12

97

97

Control menu

Control context

Parameters

Current settings

Note

If you decrease an upper alarm limit to a value almost as low as the lower limit,
the lower limit decreases so that it is always lower than the upper limit.

If you increase a lower alarm limit to a value almost as high as the upper limit, the
upper limit increases so that it is always higher than the lower limit.