Overview of monitor operation

Welch Allyn Propaq LT Vital Signs Monitor

About the Resp control menu

Using setup menus

Use the Setup menus—Alarms, ECG, NIBP, SpO2, and Timings—to define settings for
monitor behavior.

Parameter

Options

Parameter

Options

Resp Monitoring

Off On

Lower Alarm

Off On

Upper Alarm

Off On

Lower Limit

Upper Limit

Adult

2 - 148

Adult

4 - 150

Pediatric

2 - 148

Pediatric

4 - 150

Neonate

3 - 148

Neonate

5 - 150

Resp Lead

Ld1 (RA-LA) Ld2 (RA

-

LL)

Setting

Options

Alarms

Suspend Audible Alarms

Off, On (with a time value)

Alarm Tone

Low, Med, High

ECG

Resp Monitoring

Off, On

ECG Bandwidth

Monitor, Extended

Power Source Filter

60 Hz, 50 Hz, Off

Pacer Indicator

Off, On

NIBP

NIBP Format

SD, SD(m), sd(M)

NIBP Units

mmHg, kPa

Smartcuf

Off, On

SpO

2

(Masimo only)

Sensitivity

Normal, Maximum, APOD

FastSAT

Off, On

Averaging Time

2, 4, 8, 10, 12, 14, 16

Timings

Back Light Time Out (Min)

2, 5, 10, 15, 30, On, Off

Display Time Out (Min)

2, 5, 10, 15, 30, On

Demo Mode

Disabled, Low, High