background image

132

Welch Allyn Propaq LT Vital Signs Monitor

Resp On Default

Alarm Limits

HR/PR

Resp

SpO

2

NIBP

Adult

Pediatric

Neonatal

(Yes)

(Yes)

(Yes)

(No)

(No)

(No)

Adult

Pediatric

Neonatal

Upper On Off 120 ___

Upper

On Off 150 ___

Upper

On Off 200 ___

Lower On Off

50 ___

Lower On Off

50 ___

Lower

On Off 100 ___

Adult

Pediatric

Neonatal

Upper On Off

30 ___

Upper

On Off

45 ___

Upper

On Off

60 ___

Lower On Off

5 ___

Lower On Off

10 ___

Lower

On Off

10 ___

Adult

Pediatric

Neonatal

Upper On Off 100 ___

Upper

On Off 100 ___

Upper

On Off

95 ___

Lower On Off

90 ___

Lower On Off

90 ___

Lower

On Off

85 ___

Adult

Pediatric

Neonatal

Systolic

Systolic

Systolic

Upper

On Off 220 ___

Upper

On Off 145

___

Upper

On Off 100

___

Lower On Off

75 ___

Lower

On Off

75

___

Lower

On Off

50

___

Diastolic

Diastolic

Diastolic

Upper

On Off 110 ___

Upper

On Off 100

___

Upper

On Off

70

___

Lower On Off

35 ___

Lower

On Off

35

___

Lower

On Off

30

___

Mean

Mean

Mean

Upper

On Off 120 ___

Upper

On Off 110

___

Upper

On Off

80

___

Lower On Off

50 ___

Lower

On Off

50

___

Lower

On Off

35

___