Alarms and alerts
Welch Allyn Propaq LT Vital Signs Monitor
Figure 86. Sample alarm screen
Figure 87. Sample alert screen
2.
Check the patient and provide appropriate care.
3.
Press
to silence the audible tone for 90 seconds at the monitor and at Acuity.
•
Silencing the audible tone does not remove visual alarm or alert indications.
•
Silencing the audible tone causes an alarm or alert suspend at Acuity.
•
When the alarm or alert condition is corrected, all alarm or alert indicators cease
and all alarms are immediately rearmed.
4.
After caring for the patient, verify that alarm limits are enabled and correctly set.
Suspending the alarm tone
If this feature is enabled in the monitor configuration (See
page 103), you can suspend all alarm tones for all parameters—preventing the alarm tone
from sounding if an alarm condition occurs—while monitoring a patient. If an alarm
condition occurs while the alarm tones are suspended, the monitor presents visual alarm
indicators but does not sound the tone.
In the monitor configuration, the alarm tone suspension period can be set to Disabled, to
Always On, or to a period: 90 sec or 2, 3, 4, 5, 10, 15, 30, or 60 minutes.
•
If it is configured to Disabled, you cannot suspend the alarm tone at the monitor.
WARNING If you turn off or modify any alarm limits while responding to an
alarm, restore those alarm limits before you resume monitoring.
%
140/78
140/78
125
125
31
31
88
88
HALL, ROBERT E.
HALL, ROBERT E.
3456187
3456187
3:00:06P
3:00:06P
Adult
Adult
Rm 239
Rm 239
HR/min
HR/min
NIBP mmHg (102)
NIBP mmHg (102)
@2:47P Manual
@2:47P Manual
Resp/min
Resp/min
SpO2
SpO2
II 1mV/cm
II 1mV/cm
X
Resp/min
Resp/min
SpO2
SpO2
I 1mV/cm
I 1mV/cm
HALL, ROBERT E.
HALL, ROBERT E.
3456187
3456187
12:41:32
12:41:32
Adult
Adult
Rm 239
Rm 239
Audible Alarms Silenced
Audible Alarms Silenced
80
80
140/78
140/78
12
12
97
97 %
HR/min
HR/min
NIBP mmHg (102)
NIBP mmHg (102)
@2:47P Manual
@2:47P Manual
EQUIPMENT ALERT
EQUIPMENT ALERT
ECG Fault
ECG Fault
LL Lead Failed
LL Lead Failed
ACKNOWLEDGE
ACKNOWLEDGE
Note
The factory default suspension period is 4 minutes.