Feature enable, System features, Lockouts – Welch Allyn 901061 Propaq LT Vital Signs Monitor - User Manual User Manual
Page 139

Directions for Use
135
Feature Enable
System Features
Enable Wireless Communications (with Acuity)
(Yes)
(No)
For wireless monitors that must communicate with Acuity, this must be set to
(Yes). If
this is set to
(No), the monitor cannot communicate with Acuity.
For standalone monitors that do not communicate with Acuity, always set this to
(No).
If this is set to
(Yes), the configuration file is not valid and cannot be uploaded to the
monitor.
NIBP Turbo Mode Enable
(Yes)
(No)
When the monitor takes automatic NIBP measurements in Turbo Mode, it takes as many
measurements as possible within the 5-minute Turbo Mode measurement period.
SpO
2
Spot Check Enable
(Yes)
(No)
If SpO
2
Spot Check is enabled, a clinician can perform a quick SpO
2
measurement when
SpO
2
monitoring is turned off. For detailed information, see Propaq LT Vital Signs Monitor
Directions for Use.
Lockouts
If lockouts are enabled, simultaneously pressing
the Left Arrow, Right Arrow,
and Down Arrow buttons) locks out one or more of the following:
Display Lockout Enable
(Yes)
(No)
If display lockout is enabled,
locks out the monitor display until the combination
is repeated or until an alarm or alert condition arises.
Note
If
(Yes) is selected, NIBP Units Default (see
“NIBP Units Default”
on
page 127) must be set to mmHg.
Display
Prevent unauthorized viewing of confidential patient information
Back Light
Extend the battery charge life
Buttons
Block access to the monitor buttons