Directions for Use

135

Feature Enable

System Features

Enable Wireless Communications (with Acuity)

(Yes)
(No)

For wireless monitors that must communicate with Acuity, this must be set to

(Yes). If

this is set to

(No), the monitor cannot communicate with Acuity.

For standalone monitors that do not communicate with Acuity, always set this to

(No).

If this is set to

(Yes), the configuration file is not valid and cannot be uploaded to the

monitor.

NIBP Turbo Mode Enable

(Yes)
(No)

When the monitor takes automatic NIBP measurements in Turbo Mode, it takes as many
measurements as possible within the 5-minute Turbo Mode measurement period.

SpO

2

Spot Check Enable

(Yes)
(No)

If SpO

2

Spot Check is enabled, a clinician can perform a quick SpO

2

measurement when

SpO

2

monitoring is turned off. For detailed information, see Propaq LT Vital Signs Monitor

Directions for Use.

Lockouts

If lockouts are enabled, simultaneously pressing

the Left Arrow, Right Arrow,

and Down Arrow buttons) locks out one or more of the following:

Display Lockout Enable

(Yes)
(No)

If display lockout is enabled,

locks out the monitor display until the combination

is repeated or until an alarm or alert condition arises.

Note

If

(Yes) is selected, NIBP Units Default (see

“NIBP Units Default”

on

page 127) must be set to mmHg.

Display

Prevent unauthorized viewing of confidential patient information

Back Light

Extend the battery charge life

Buttons

Block access to the monitor buttons