Mode settings, Smartcuf as default for nibp – Welch Allyn 901061 Propaq LT Vital Signs Monitor - User Manual User Manual

Page 135

background image

Directions for Use

131

Mode Settings

Default Resp Lead

NIBP Mean Numerics

NIBP Mode

NIBP Auto Mode Interval

Use Smartcuf as Default for NIBP

An NIBP measurement can be adversely affected by many factors, including body motion
(such as convulsions or shivering), ambient vibration, vehicle motion, weak pulse, a
sudden change in blood pressure, or sudden cuff movement. When these factors are
present, Welch Allyn’s patented Smartcuf technology can greatly increase the accuracy of
NIBP measurements.

Adult

Pediatric

Neonatal

__ Lead 1 (RA-LA)

__ Lead 1 (RA-LA)

__ Lead 1 (RA-LA)

__ Lead 2 (RA-LL)

__ Lead 2 (RA-LL)

__ Lead 2 (RA-LL)

Adult

Pediatric

Neonatal

Small

Small

Small

Large

Large

Large

Off

Off

Off

Note

Selecting Large reduces the size of the systolic and diastolic displays.

Adult

Pediatric

Neonatal

Auto, Manual

Auto, Manual

Auto, Manual

Adult (minutes)

Pediatric (minutes)

Neonatal (minutes)

1, 2, 3, 5, 10, 15, 30, 60

1, 2, 3, 5, 10, 15, 30, 60

1, 2, 3, 5, 10, 15, 30, 60

Adult

Pediatric

Neonatal

(Yes)

(Yes)

(Yes)

(No)

(No)

(No)

Note

Smartcuf functions only during ECG monitoring.