Welch Allyn 901061 Propaq LT Vital Signs Monitor - User Manual User Manual
Page 177

Directions for Use
Compliance
173
Guidance and manufacturer’s declaration—electromagnetic immunity
The Propaq LT Series monitor with Charging/Communication Cradle and Large Color Display Interface is intended for use
in the electromagnetic environment specified below. The customer or the user of the Propaq LT Series monitor with
Charging/Communication Cradle and Large Color Display Interface should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment—guidance
Portable and mobile RF communications equipment should
be used no closer to any part of the Propaq LT Series monitor
with Charging/Communication Cradle and Large Color
Display Interface, including cables, than the recommended
separation distance calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
Conducted RF
IEC 61000
-
4
-
6
3 V
rms
150 kHz to 80 MHz
2Hz AM
3 V
rms
d = 1.2
Radiated RF
IEC 61000
-
4
-
3
3 V/m
80 MHz to 2.5 GHz
2Hz AM
3 V/m
d = 1.2
80 MHz to 800 MHz
d = 2.3
800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts according to the transmitter
manufacturer and d is the recommended separation distance
in meters.
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey
a
, should be less than the
compliance level in each frequency range
b
.
Interference may occur in the vicinity of equipment marked
with the following symbol:
Note 1
At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the Propaq LT Series monitor with Charging/Communication Cradle and Large Color Display
Interface is used exceeds the applicable RF compliance level above, the Propaq LT Series monitor with Charging/Communication
Cradle and Large Color Display Interface should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the Propaq LT Series monitor with
Charging/Communication Cradle and Large Color Display Interface.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
P
P
P