Welch Allyn 901061 Propaq LT Vital Signs Monitor - User Manual User Manual
Page 179

Directions for Use
Compliance
175
Monitor, battery-operated, with cable 008-0799-00/01
The EMC specifications listed on pages
to
apply only to Propaq LT monitors using
ECG cables 008-0799-00 and 008-0799-01. For general EMC specifications, see
Guidance and manufacturer’s declaration—electromagnetic emissions
The battery-operated Propaq LT Series monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of the monitor should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment—guidance
RF emissions
CISPR 11
Group 1
The battery-operated Propaq LT Series monitor uses RF
energy only for its internal function.
a
Therefore, its RF
emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
a.
The battery-operated Propaq LT Series monitor contains either a 5-GHz orthogonal frequency-division multiplexing transmitter or a
2.4
-
GHz frequency
-
hopping spread
-
spectrum transmitter for the purpose of wireless communication. The radio is operated according
to the requirements of various agencies, including FCC 47 CFR 15.247 and R&TTE Directive (1995/5/EC). The radio is excluded from
the EMC requirements of 60601
-
1
-
2:2001, but should be considered when addressing possible interference issues between this and
other devices.
RF emissions
CISPR 11
Class B
The battery-operated Propaq LT Series monitor is suitable for
use in all establishments, including domestic establishments
and those directly connected to the public low
-
voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000
-
3
-
2
No connection to mains
(battery
-
operated)
Voltage fluctuations/flicker
emissions
IEC 61000
-
3
-
3
No connection to mains
(battery
-