ParamSet can be used multiple times for any alarm limit. Each time you press ParamSet
for a given vital sign, the alarm limit expands by the configured percentage. However, the
alarm limits can never be expanded beyond the limit boundaries built into the monitor.
(See the specifications for vital signs in Appendix A, starting with

“ECG”

on page 143.)

This table shows the effect of ParamSet for various vital signs, based on a patient’s
normal levels for each vital sign. These values are based on the assumption that:

the patient mode is Adult

the monitor is initially using factory default alarm levels

For each parameter listed below, the factory value is shown in bold. The factory value is
used until another value is uploaded to the monitor.

Note

The ParamSet default values can be changed only by downloading another
monitor configuration.

When monitor power is cycled, the factory default alarm limits and ParamSet
default values are again in effect.

Parameter

Alarming

Level

Alarm Limit

(ParamSet

= 5%)

Alarm Limit

(Paramset

= 10%)

Alarm Limit

(ParamSet

= 15%)

Alarm Limit

(Paramset

= 20%)

Alarm Limit

(Paramset

= 25%)

HR/PR

Upper

90

94

99

104

108

113

Lower

60

57

54

51

48

45

NIBP Systolic Upper

140

147

154

161

168

175

Lower

100

95

90

85

80

75

NIBP
Diastolic

Upper

90

94

99

104

108

113

Lower

60

57

54

51

48

45

NIBP MAP

Upper

107

112

118

123

128

134

Lower

73

69

66

62

58

55

SpO

2

Upper

100

100

100

100

100

100

Lower

90

86

81

77

72

68

Resp

Upper

20

21

22

23

24

25

Lower

12

11

11

10

10

9