Welch Allyn 901061 Propaq LT Vital Signs Monitor - User Manual User Manual
Page 138

134
Welch Allyn Propaq LT Vital Signs Monitor
ParamSet can be used multiple times for any alarm limit. Each time you press ParamSet
for a given vital sign, the alarm limit expands by the configured percentage. However, the
alarm limits can never be expanded beyond the limit boundaries built into the monitor.
(See the specifications for vital signs in Appendix A, starting with
This table shows the effect of ParamSet for various vital signs, based on a patient’s
normal levels for each vital sign. These values are based on the assumption that:
•
the patient mode is Adult
•
the monitor is initially using factory default alarm levels
For each parameter listed below, the factory value is shown in bold. The factory value is
used until another value is uploaded to the monitor.
Note
The ParamSet default values can be changed only by downloading another
monitor configuration.
When monitor power is cycled, the factory default alarm limits and ParamSet
default values are again in effect.
Parameter
Alarming
Level
Alarm Limit
(ParamSet
= 5%)
Alarm Limit
(Paramset
= 10%)
Alarm Limit
(ParamSet
= 15%)
Alarm Limit
(Paramset
= 20%)
Alarm Limit
(Paramset
= 25%)
HR/PR
Upper
90
94
99
104
108
113
Lower
60
57
54
51
48
45
NIBP Systolic Upper
140
147
154
161
168
175
Lower
100
95
90
85
80
75
NIBP
Diastolic
Upper
90
94
99
104
108
113
Lower
60
57
54
51
48
45
NIBP MAP
Upper
107
112
118
123
128
134
Lower
73
69
66
62
58
55
SpO
2
Upper
100
100
100
100
100
100
Lower
90
86
81
77
72
68
Resp
Upper
20
21
22
23
24
25
Lower
12
11
11
10
10
9