Directions for Use

Monitor configuration

111

SpO

2

pediatric

Upper Alarm Limit

100%

52% - 100%

Monitor, Acuity

Lower Alarm Limit

90%

50% - 98%

Monitor, Acuity

SpO

2

neonatal

Upper Alarm Limit

95%

52% - 100%

Monitor, Acuity

Lower Alarm Limit

85%

50% - 98%

Monitor, Acuity

SpO

2

Masimo

Perfusion Index

Enabled

Enabled, Not enabled

Configuration Utility

Signal IQ

Enabled

Enabled, Not enabled

Configuration Utility

Resp general

ParamSet for Upper Resp

5%

5% - 25%

Configuration Utility

ParamSet for Lower Resp

5%

5% - 25%

Configuration Utility

Resp adult

Resp On/Off

No

Yes, No

Monitor, Configuration
Utility

Vector

Ld1

Ld1 (RA-LA), Ld2 (RA-LL)

Monitor, Configuration
Utility, Acuity

Upper Alarm Limit

30/min

4/min - 150/min

Monitor, Acuity

Lower Alarm Limit

5/min

2/min - 148/min

Monitor, Acuity

Resp pediatric

Resp On/Off

Yes

Yes, No

Monitor, Configuration
Utility

Vector

Ld1

Ld1 (RA-LA), Ld2 (RA-LL)

Monitor, Configuration
Utility, Acuity

Upper Alarm Limit

45/min

4/min - 150/min

Monitor, Acuity

Lower Alarm Limit

10/min

2/min - 148/min

Monitor, Acuity

Resp neonatal

Resp On/Off

Yes

Yes, No

Monitor, Configuration
Utility

Vector

Ld1

Ld1 (RA-LA), Ld2 (RA-LL)

Monitor, Configuration
Utility, Acuity

Upper Alarm Limit

60/min

5/min - 150/min

Monitor, Acuity

Lower Alarm Limit

10/min

3/min - 148/min

Monitor, Acuity

Table 10. Parameter configuration matrix (continued)

Parameter

Factory setting

Possible values

Where set