Directions for Use

Standalone monitoring

63

Figure 68. SpO

2

monitoring turned off

To take an SpO

2

spot check reading

1.

Enable spot checks. (See

“To prepare to take a spot check reading”

on page 62.)

2.

Attach the sensor to the monitor and the patient.

3.

Highlight SpO2 @ XX:XX and press

. The SpO

2

drop-down menu appears.

Figure 69. SpO

2

drop-down menu

4.

Press

to highlight Spot Check.

The drop-down menu disappears. SEARCH appears above SpO

2

Spot Check (with

pulse rate if SpO

2

is the source of pulse rate).

Figure 70. SpO

2

spot check: waiting for an SpO

2

signal

After a few seconds, the SpO

2

heart-beat indicator starts showing heart beats.

After about 30 seconds, SEARCH disappears and the pulse oximetry reading
appears.

Note

The SpO

2

drop-down menu can be accessed only when SpO

2

is set to Off.

80

80

HALL, ROBERT E.

HALL, ROBERT E.

3456187

3456187

3:11:24P

3:11:24P

Adult

Adult

Rm 239

Rm 239

NIBP mmHg

NIBP mmHg

Resp/min

Resp/min

HR/min

HR/min

SpO2 @ 3:03P

SpO2 @ 3:03P

%

S

D

M

SpO

2

Monitoring is Off

SpO2

SpO2

On

On
Off

Off
Spot Check

Spot Check

Enable continuous automatic SpO

2

function

Disable automatic SpO

2

function (enable Spot Checks)

Take a Spot Check reading

HALL, ROBERT E.

HALL, ROBERT E.

3456187

3456187

3:11:30P

3:11:30P

Adult

Adult

Rm 239

Rm 239

Resp/min

Resp/min

PR/min

PR/min

SpO2 Spot Check

SpO2 Spot Check

%

NIBP mmHg

NIBP mmHg

SEARCH

SEARCH

S

D

M

SEARCH indicates that the monitor
is waiting for pulse oximetry data.