Spo2, Spo2 on on off off spot check spot check – Welch Allyn 901061 Propaq LT Vital Signs Monitor - User Manual User Manual
Page 67

Directions for Use
Standalone monitoring
63
Figure 68. SpO
2
monitoring turned off
To take an SpO
2
spot check reading
1.
Enable spot checks. (See
“To prepare to take a spot check reading”
2.
Attach the sensor to the monitor and the patient.
3.
Highlight SpO2 @ XX:XX and press
. The SpO
2
drop-down menu appears.
Figure 69. SpO
2
drop-down menu
4.
Press
to highlight Spot Check.
•
The drop-down menu disappears. SEARCH appears above SpO
2
Spot Check (with
pulse rate if SpO
2
is the source of pulse rate).
Figure 70. SpO
2
spot check: waiting for an SpO
2
signal
•
After a few seconds, the SpO
2
heart-beat indicator starts showing heart beats.
•
After about 30 seconds, SEARCH disappears and the pulse oximetry reading
appears.
Note
The SpO
2
drop-down menu can be accessed only when SpO
2
is set to Off.
80
80
HALL, ROBERT E.
HALL, ROBERT E.
3456187
3456187
3:11:24P
3:11:24P
Adult
Adult
Rm 239
Rm 239
NIBP mmHg
NIBP mmHg
Resp/min
Resp/min
HR/min
HR/min
SpO2 @ 3:03P
SpO2 @ 3:03P
%
S
D
M
SpO
2
Monitoring is Off
SpO2
SpO2
On
On
Off
Off
Spot Check
Spot Check
Enable continuous automatic SpO
2
function
Disable automatic SpO
2
function (enable Spot Checks)
Take a Spot Check reading
HALL, ROBERT E.
HALL, ROBERT E.
3456187
3456187
3:11:30P
3:11:30P
Adult
Adult
Rm 239
Rm 239
Resp/min
Resp/min
PR/min
PR/min
SpO2 Spot Check
SpO2 Spot Check
%
NIBP mmHg
NIBP mmHg
SEARCH
SEARCH
S
D
M
SEARCH indicates that the monitor
is waiting for pulse oximetry data.