Time-outs – Welch Allyn 901061 Propaq LT Vital Signs Monitor - User Manual User Manual
Page 140

136
Welch Allyn Propaq LT Vital Signs Monitor
Back Light Lockout Enable
(Yes)
(No)
If back light lockout is enabled, the 3-button combination
locks out the monitor
display back light until the combination is repeated or until an alarm or alert condition
arises.
Buttons Lockout Enable
(Yes)
(No)
If you enable buttons lockout, the 3-button combination
locks out normal button
operation until the combination is repeated or until an alarm or alert condition arises.
Time-Outs
Display Time-Out
__ 2 minutes
__ 5 minutes
__ 10 minutes
__ 15 minutes
__ 30 minutes
__ Always On (display always on)
Display Time Out greatly increases monitor run time on battery power. Display Time Out
completely shuts off the display if the monitor detects no button press and no alarm for a
number of minutes defined below. When the display is off, no monitor information is
visible.
Any alarm condition, alert condition, or button press immediately restores the display. If
the buttons are locked out (see
“Buttons Lockout Enable”
on page 136), pressing
and then any other button immediately restores the display.
Back Light Time-Out
__ Always Off
__ 2 minutes
__ 5 minutes
__ 10 minutes
__ 15 minutes
__ 30 minutes
__ Always On (back light always on)
Back Light Time Out increases monitor run time on battery power. Back Light Time Out
turns off the back light if the monitor detects no button press and no alarm for the number
of minutes defined below. When the back light is off, the display continues to be visible,
although slightly dimmer, in ambient light.
Any alarm condition, alert condition, or button press immediately restores the back light. If
the buttons are locked out (see
“Buttons Lockout Enable”
on page 136), pressing
and then any other button immediately restores the back light.