beautypg.com

Memory settings – Teledyne LeCroy Sierra M124 User Manual User Manual

Page 100

background image

Teledyne LeCroy

SAS/SATA Software Menus and Toolbars

98

 

 

Sierra M124 SAS/SATA Protocol Analyzer User Manual

Figure 2.69: SAS: Setting Project Options

SAS vs. SATA: For the ALIGN Transmission Period section, SATA Dialog shows options 256 
and 258, and does not show 2048 or 2049.
For Speed, SATA Dialog shows H1, D1 to H4, D4 and does not show I1, T1 to I4, T4.
SATA Dialog does not show MUX Setting button.

SATA dialog shows DevSlp min detection time field to enter the debouncer value.

Memory Settings

The Trace Memory Status section has the following fields.

Trigger Position

Pre‐Trigger defaults to 50%, which defines the percentage of data to capture before and 
after the triggering event. You can change this percentage by dragging the slider.

Capture of the specified percentage of the data prior to the triggering event cannot be 
guaranteed and may be 0. This can occur if the triggering event occurs before storing the 
required amount of pre‐trigger event data. In such a case, the data display shows fewer 
than the specified data points prior to the triggering event. For more detail, (see 

“Pre‐

Trigger” on page 97.

)

Note:

Trigger Position only works when the triggering option is Pattern.