beautypg.com

Creating a calibration table using the legacy 6000 – TeeJet Legacy 6000 Manual User Manual

Page 156

background image

Legacy 6000

98-05053 R2

151

1. enter Channel Setup and navigate to Primary Sensor. Select the Settings

softkey.

2. use the Arrow keys to highlight Cal# Basis and press the Enter button.
3. Select the Cal table to be used. the Name of the Cal table is the Cal

Table Identifier (the first column in the brackets when creating the Cal
table). after the Cal# Basis has been selected, continue through Channel
Setup and save to

Apply the settings. the Cal table is now ready for

use.

4. to use the Cal table enter Product setup from the ARM Launcher page.
5. use the Arrow keys to highlight the Cal Table Identifier (e.g., GATE

height) and press the Enter button. Select the desired calibration de-
scription. this selection will automatically change the Cal# in the sys-
tem.

Creating a Calibration Table using the Legacy 6000

From Channel Setup, navigate to the

Primary Sensor page and press the

Settings softkey.
use the

Arrow keys highlight Cal# Basis and press the Enter button.

Name the Cal# Basis ( the name of the Cal table). once a name has

been established, continue through Channel Setup and apply settings.
advance to the

ARM Launcher page and select the Calibration softkey.

Choose the calibration procedure for the application and press the For-
ward Arrow
softkey.

Enter button

Arrow keys

Settings softkey

Enter button

Arrow keys

Forward Arrow

softkey

Calibration

softkey

Settings softkey