VOLTage:TRIGgered

Phase Selectable

This command selects the ac rms amplitude that the output waveform will be set to during a
triggered step or pulse transient.

The maximum peak voltage that the AC source can output is 425 V peak. This includes any
combination of voltage, and function shape values. Therefore, the maximum value that can be
programmed depends on the peak-to-rms ratio of the selected waveform. For a sinewave, the
maximum voltage that can be programmed is 300 V rms.

Note: You cannot program a voltage that produces a higher volt-second on the output than a

300V rms sinewave.

Command Syntax

[SOURce:]VOLTage[:LEVel]:TRIGgered :AMPLitude]

Parameters

0 to 300 (for sinewaves)

Unit

V (rms voltage)

*RST Value

0 volt

Examples

VOLT:TRIG 120 VOLT:LEV:TRIG 120

Query Syntax

SOURce:]VOLTage[:LEVel]:TRIGgered:AMPLitude]?

Returned Parameters

If the TRIG level is not programmed, the IMM level is

returned.

Related Commands

VOLT VOLT:MODE

FUNC:SHAP

VOLTage:MODE

Phase Selectable

This command determines how the ac rms output voltage is controlled during a triggered output
transient. The choices are:

FIXed

The voltage is unaffected by a triggered output transient.

STEP

The voltage is programmed to the value set by VOLTage:TRIGgered
when a triggered transient occurs.

PULSe

The voltage is changed to the value set by VOLTage:TRIGgered for a
duration determined by the pulse commands.

LIST

The voltage is controlled by the voltage list when a triggered transient
occurs.

Command Syntax

[SOURce:]VOLTage:MODE

Parameters

FIXed | STEP | PULSe | LIST

*RST Value

FIX

Examples

VOLT:MODE LIST

VOLT:MODE FIX

Query Syntax

[SOURce:]VOLTage:MODE?

Returned Parameters

Related Commands

VOLT:TRG

VOLT