Programming Manual

Lx \ Ls Series II

92

VOLTage:ALC:SOURce

These commands select the source from which the output voltage is sensed. The following
voltage sense sources can be selected:

INTernal

This senses the voltage at the output of the power amplifier on the
inboard side of the output disconnect relay.

EXTernal

This senses the output voltage at the user's sense terminals, which
allows remote voltage sensing at the load.

Command Syntax

[SOURce:]VOLTage:ALC:SOURce

Parameters

INTernal | EXTernal

*RST Value

INTernal

Examples

VOLT:ALC:SOUR EXT

Query Syntax

[SOURce:]VOLTage:ALC:SOURce?

Returned Parameters

Related Commands

VOLT:SENS:DET

Note: The VOLT:ALC:SOUR command is an alias for the VOLT:SENS:SOUR command. Both
perform the same function. This is done for backward compatability with the Agilent 6834B.

VOLTage

Phase Selectable

This command programs the ac rms output voltage level of the AC source. The maximum peak
voltage that the AC source can output is 425 V peak. This includes any combination of voltage
and function shape values. Therefore, the maximum value that can be programmed depends on
the peak-to-rms ratio of the selected waveform. For a sinewave, the maximum voltage that can be
programmed is 300 V rms.

Note: You cannot program a voltage that produces a higher volt-second on the output than a

300V rms sinewave.

Command Syntax

[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]

Parameters

0 to 300 (for sinewaves)

Unit

V (rms voltage)

*RST Value

1 volt

Examples

VOLT 250

VOLT:LEV 25

Query Syntax

[SOURce:]VOLTage[:LEVel]

[:IMMediate][:AMPLitude]?

Returned Parameters

Related Commands

VOLT:MODE

VOLT:TRIG

FUNC:SHAP