AMETEK Lx Series II Programming Manual User Manual

Page 76

background image

Programming Manual

Lx \ Ls Series II

72

CURRent:PROTection:STATe

This command enables or disables the AC source overcurrent (OC) protection function. If the
overcurrent protection function is enabled and the AC source exceeds the programmed level, then
the output is disabled and the Questionable Condition status register OC bit is set (see Chapter
7). An overcurrent condition can be cleared with OUTPut:PROTection:CLEar after the cause of
the condition is removed.

Use OUTP:PROT:DEL to prevent momentary current limit conditions caused by programmed
output changes from tripping the over current protection. Use CURR:PROT:DEL to hold off
tripping the output due to temporary overload conditions.

Command Syntax

[SOURce:]CURRent:PROTection:STATe

Parameters

0 | 1 | OFF | ON

*RST Value

OFF

Examples

CURR:PROT:STAT 0 CURR:PROT:STAT OFF

Query Syntax

[SOURce:]CURRent:PROTection:STATe?

Returned Parameters

0 | 1

Related Commands

OUTP:PROT:CLE

CURR:PROT:DEL