Voltage measurement subsystem – AMETEK Lx Series II Programming Manual User Manual
Page 63
Programming Manual
Lx \ Ls Series II
59
4.12 Voltage Measurement Subsystem
This subsystem programs the voltage measurement capability of the Lx\Ls Series. Two
measurement commands are available: MEASure and FETCh. MEASure triggers the acquisition
of new measurement data before returning a reading. FETCh returns a reading computed from
previously acquired data.
Individual outputs of a three-phase source are specified by the setting of INSTrument:NSELect.
Subsystem Syntax
MEASure | FETCh
[:SCALar]
:VOLTage
[:DC]?
Returns the dc component of the voltage
:AC?
Returns ac rms voltage
:ACDC?
Returns ac+dc rms voltage
:HARMonic
[:AMPLitude]?
Returns amplitude of the Nth harmonic of voltage
:PHASe?
Returns phase of the Nth harmonic of voltage
:THD?
Returns % of total harmonic distortion of voltage
:THD:MODE RMS | FUND
Sets THD calculation to either RMS or Fundamental
mode.
MEASure:VOLTage?
FETCh:VOLTage?
Phase Selectable
These queries return the dc component of the output voltage being sourced at the output
terminals.
Query Syntax
MEASure[:SCALar]:VOLTage[:DC]?
FETCh[:SCALar]:VOLTage[:DC]?
Parameters
None
Examples
MEAS:VOLT? FETC:VOLT?
Returned Parameters
Related Commands
INST:NSEL
MEASure:VOLTage:AC?
FETCh:VOLTage:AC?
Phase Selectable
These queries return the ac rms voltage being sourced at the output terminals.
Query Syntax
MEASure[:SCALar]:VOLTage:AC?
FETCh[:SCALar]:VOLTage:AC?
Parameters
None
Examples
MEAS:VOLT:AC?
FETC:VOLT:AC?
Returned Parameters
Related Commands
INST:NSEL
MEASure:VOLTage:ACDC?
FETCh:VOLTage:ACDC?
Phase Selectable
These queries return the ac or dc rms voltage being sourced at the output terminals.