B and C (ø1, ø2 and ø3). If the MODE field is set (standard on Lx Series,
optional on Ls Series) the AC source can operate in both 1 and 3 phase
modes.

Other

Any value other than 0 or 120 indicates 2 phase configuration with phase
angle between A and B set to value shown.

Query Syntax

LIMit:PHASe?

Returned Parameters

Examples

LIM:PHAS?

Related Commands

SYST:CONF:NOUT?

LIMit:VOLTage:HIGH

Query form returns the maximum available output voltage for the high voltage range. This value
determines the maximum AC RMS voltage available using a sinusoidal waveform. Note that this
is not the same as the available maximum voltage, which is a function crest factor of the voltage
waveform. To query the maximum available voltage, use the VOLT? MAX command.

Query Syntax

LIMit:VOLTage:HIGH?

Returned Parameters

Examples

LIM:VOLT:HIGH?

Related Commands

LIM:VOLT:LOW?

LIMit:VOLTage:LOW

Query form returns the maximum available output voltage for the low voltage range. This value
determines the maximum AC RMS voltage available using a sinusoidal waveform. Note that this
is not the same as the available maximum voltage, which is a function crest factor of the voltage
waveform. To query the maximum available voltage, use the VOLT? MAX command.

Query Syntax

LIMit:VOLTage:LOW?

Returned Parameters

Examples

LIM:VOLT:LOW?

Related Commands

LIM:VOLT:HIGH?