AMETEK Lx Series II Programming Manual User Manual
Page 121
![background image](https://www.manualsdir.com/files/771680/content/doc121.png)
Programming Manual
Lx \ Ls Series II
117
Command Syntax
TRIGger:SEQuence3[:IMMediate]
TRIGger:ACQuire[:IMMediate]
Parameters
None
Examples
TRIG:ACQ
TRIG:SEQ3:IMM
Related Commands
ABOR TRIG:SOUR
TRIG:DEL
TRIG:SYNC
TRIG:SYNC:PHAS
INIT
INIT:CONT
*TRG *WAI
TRIGger:SEQuence3:SOURce
TRIGger:ACQuire:SOURce
These commands select the trigger source for a triggered measurement sequence as follows:
BUS
IEEE-488 device, *TRG, or
EXTernal
backpanel Trigger In1 SMA connector
TTLTrg
the signal driving the Trigger Out1 SMA
Command Syntax
TRIGger:SEQuence3:SOURce
Command Syntax
TRIGger:ACQuire:SOURce
Parameters
BUS | EXT | TTLT
*RST Value
BUS
Examples
TRIG:ACQ:SOUR BUS
TRIG:SEQ3:SOUR EXT
Query Syntax
TRIGger:SEQuence3:SOURce?
TRIGger:ACQuire:SOURce?
Returned Parameters
Related Commands
ABOR TRIG:SOUR
TRIG:DEL
TRIG:SYNC
TRIG:SYNC:PHAS
INIT
INIT:CONT
*TRG *WAI
TRIGger:SEQuence1:DEFine
TRIGger:SEQuence2:DEFine
TRIGger:SEQuence3:DEFine
These commands define the names that are aliased to trigger sequences 1, 2 and 3. The
command accepts only TRANsient for sequence 1, SYNChronous for sequence 2, and ACQuire
for sequence 3 as predefined names. The query allows the user to query the instrument names
aliased to sequences 1, 2, and 3.
Command Syntax
TRIGger:SEQuence1:DEFine TRANsient
TRIGger:SEQuence2:DEFine SYNChronous
TRIGger:SEQuence3:DEFine ACQuire
Parameters
TRANsient, SYNChronous, ACQuire
Examples
SEQ1:DEF TRAN SEQ3:DEF ACQ
Query Syntax
TRIGger:SEQuence1:DEFine?
TRIGger:SEQuence2:DEFine?
TRIGger:SEQuence3:DEFine?
Returned Parameters