4.21 Source Subsystem - Pulse
This subsystem controls the generation of output pulses. The PULSe:DCYCle, PULSe:HOLD,
PULSe:PERiod, and PULSe:WIDTh commands are coupled, which means that the values
programmed by any one of these commands can be affected by the settings of the others. Refer
to the tables under PULSe:HOLD for an explanation of how these commands affect each other.
Subsystem Syntax
[SOURce:]
PULSe
:COUNt
Selects transient pulse count
:DCYCle
Selects pulse duty cycle
:HOLD WIDTh |DCYCle
Selects parameter that is held constant
:PERiod
Selects pulse period when the count is greater than 1
:WIDTh
Selects width of the pulses
PULSe:COUNt
This command sets the number of pulses that are output when a triggered output transient
occurs. The command accepts parameters in the range 1 through 2E8. If INFinity or MAXimum is
sent, the output pulse repeats indefinitely.
Command Syntax
[SOURce:]PULSe:COUNt
Parameters
1 to 2E8 | MINimum | MAXimum | INFinity
*RST Value
1
Examples
PULS:COUN 3 PULS:COUN MIN PULS:COUN INF
Query Syntax
[SOURce:]PULS:COUNt?
Returned Parameters
Related Commands
PULS:DCYC PULS:HOLD PULS:PER PULS:WIDT
PULSe:DCYCle
This command sets the duty cycle of the triggered output pulse. The duty cycle units are specified
in percent.
Command Syntax
[SOURce:]PULSe:DCYCle
Parameters
0 to 100%|MINimum|MAXimum
*RST Value
50%
Examples
PULS:DCYC 75 PULS:DCYC MAX
Query Syntax
[SOURce:]PULSe:DCYCle?
Returned Parameters
Related Commands
PULS:COUN PULS:HOLD PULS:PER PULS:WIDT
PULSe:HOLD
This command specifies whether the pulse width or the duty cycle is to be held constant when the
pulse period changes. The following tables describe how the duty cycle, period, and width are
affected when one, two, or all three parameters are set in a single program message.
Command Syntax
[SOURce:]PULSe:HOLD
Parameters
WIDTh|DCYCle
*RST Value
WIDTh
Examples
PULS:HOLD DCYC
Query Syntax
[SOURce:]PULSe:HOLD?
Returned Parameters
Related Commands
PULS:COUN PULS:DCYC PULS:PER PULS:WIDT