4.3

Calibration Subsystem

The commands in this subsystem allow you to do the following:

Enter the calibration password

Calibrate the current and voltage output levels, and store new calibration constants in
nonvolatile memory.

Subsystem Syntax

CALibrate

:MEASure

:CURRent

Begin current measurement calibration sequence

:SPHase

Query format returns single phase current mea coefficient.

:VOLTage

Begin current measurement calibration sequence

:PASSword Unlock calibration

:SAVE

Save new cal constants in non-volatile memory

[:SOURce]

PHASe

Phase offset calibration

:VOLTage:HRANge

Query output voltage cal coefficient

CALibrate:MEASure:CURRent

Phase Selectable

This command can only be used in calibration mode. It initiates the calibration of the ac current
metering circuits and sets the reference value to calibrate to. The query format returns the actual
calibration coefficient. Use the INST:SEL or INST:NSEL to select the desired phase.

Command Syntax

CALibrate:MEASure:CURRent

Parameters

Reference instrument current reading

Query Syntax

CALibrate:MEASure:CURRent?

Returned Parameters

Examples

CAL:MEAS:CURR 12.23

Related Commands

CAL:SAVE

CAL:MEAS:VOLT

CALibrate:MEASure:CURRent:SPHase

This command can only be used in calibration mode and in query form. It applies only to phase 1
(Lx) or A (Ls). It returns the single-phase mode current measurement calibration coefficient for Lx
models and Ls models that have the

–MODE option. Its purpose is to be able to read the

calibration status without having to switch the AC source to single phase mode.

If the Lx or Ls is in single phase mode, this command is equivalent to the the CAL:MEAS:CURR?
query for phase 1/A except it can not be used to perform a calibration. It does not initiate a
calibration. Use the CAL:MEAS:CURR? command to do this instead.

Query Syntax

CALibrate:MEASure:CURRent:SPHase?

Returned Parameters

Examples

CAL:MEAS:CURR:SPH?

Related Commands

CAL:SAVE

CAL:MEAS:VOLT

CALibrate:MEASure:VOLTage

Phase Selectable

This command can only be used in calibration mode. It initiates the calibration of the ac voltage
metering circuits. The query format returns the actual calibration coefficient. Use the INST:SEL or
INST:NSEL to select the desired phase.