Lx \ Ls Series II
89
Parameter Set
Action
DCYCle
PERiod
WIDTh
Sets WIDTh. If WIDTh < PERiod, recalculates DCYCle;
otherwise, recalculates the PERiod and DCYCle.
Sets PERiod. If WIDTh < PERiod, recalculates DCYCle;
otherwise, recalculates the PERiod and DCYCle.
Sets WIDTh. If WIDTh < PERiod, sets the PERiod and
recalculates DCYCle; otherwise, recalculates the PERiod
and DCYCle
Sets DCYCle and recalculates PERiod
Sets DCYCle and WIDTh and recalculates PERiod
Sets DCYCle and PERiod and recalculates WIDTh
Sets WIDTh. If WIDTh < PERiod, sets the PERiod and
recalculates DCYCle; otherwise, recalculates the PERiod
and DCYCle
Table 4-1: PULSe:HOLD = WIDTh parameters
Parameter Set
Action
DCYCle
PERiod
WIDTh
Sets WIDTh and recalculates the PERiod
Sets PERiod and recalculates the WIDTh
Sets WIDTh. If WIDTh < PERiod, sets the PERiod and
recalculates DCYCle; otherwise, recalculates the PERiod
and DCYCle
Sets DCYCle and recalculates PERiod
Sets DCYCle and WIDTh and recalculates PERiod
Sets DCYCle and PERiod and recalculates WIDTh
Sets WIDTh. If WIDTh < PERiod, sets the PERiod and
recalculates DCYCle; otherwise, recalculates the PERiod
and DCYCle
Table 4-2: PULSe:HOLD = DCYCle parameters
PULSe:PERiod
This command sets the period of a triggered output transient The command parameters are
model-dependent.
Command Syntax
[SOURce:]PULSe:PERiod
Parameters
3-phase models: 0 to 1.07533E6 | MINimum | MAXimum
1-phase models: 0 to 4.30133E5 | MINimum | MAXimum
Unit
s (seconds)
*RST Value
0.03333
Examples
PER 0.001 PER MIN
Query Syntax
[SOURce:]PERiod?
Returned Parameters
Related Commands
PULS:COUN PULS:DCYC PULS:HOLD PULS:WIDT
PULSe:WIDTh