VOLTage:SENSe:SOURce
These commands select the source from which the output voltage is sensed. The following
voltage sense sources can be selected:
INTernal
This senses the voltage at the output of the power amplifier on the
inboard side of the output disconnect relay.
EXTernal
This senses the output voltage at the user's sense terminals, which
allows remote voltage sensing at the load.
Command Syntax
[SOURce:]VOLTage:SENSe:SOURce
[SOURce:]VOLTage:ALC:SOURce
Parameters
INTernal | EXTernal
*RST Value
INTernal
Examples
VOLT:SENS:SOUR INT
Query Syntax
[SOURce:]VOLTage:SENSe:SOURce?
Returned Parameters
Related Commands
VOLT:ALC:SOUR EXT
Note: The VOLT:ALC:SOUR command is an alias for the VOLT:SENS:SOUR command. Both
perform the same function. This is done for backward compatability with the Agilent 6834B.
VOLTage:SLEW
This command sets the slew rate for all programmed changes in the ac rms output voltage level
of the AC source. A parameter of MAXimum or INFinity will set the slew to its maximum possible
rate. The SCPI representation for INFinity is 9.9E37. This command does not affect the rate at
which programmed dc offset changes occur.
Command Syntax
[SOURce:]VOLTage:SLEW[:IMMediate]
Parameters
1E-3 to 9.9E37 | INFinity |MINimum | MAXimum
Unit
V/S (volts per second)
*RST Value
INFinity
Examples
VOLT:SLEW 1 VOLT:SLEW MAX
VOLT:SLEW INF
Query Syntax
[SOURce:]VOLTage:SLEW[:IMMediate]?
Returned Parameters
Related Commands
VOLT:SLEW:MODE
VOLT:SLEW:TRIG