VOLTage:SENSe:SOURce

These commands select the source from which the output voltage is sensed. The following
voltage sense sources can be selected:

INTernal

This senses the voltage at the output of the power amplifier on the
inboard side of the output disconnect relay.

EXTernal

This senses the output voltage at the user's sense terminals, which
allows remote voltage sensing at the load.

Command Syntax

[SOURce:]VOLTage:SENSe:SOURce

[SOURce:]VOLTage:ALC:SOURce

Parameters

INTernal | EXTernal

*RST Value

INTernal

Examples

VOLT:SENS:SOUR INT

Query Syntax

[SOURce:]VOLTage:SENSe:SOURce?

Returned Parameters

Related Commands

VOLT:ALC:SOUR EXT

Note: The VOLT:ALC:SOUR command is an alias for the VOLT:SENS:SOUR command. Both
perform the same function. This is done for backward compatability with the Agilent 6834B.

VOLTage:SLEW

This command sets the slew rate for all programmed changes in the ac rms output voltage level
of the AC source. A parameter of MAXimum or INFinity will set the slew to its maximum possible
rate. The SCPI representation for INFinity is 9.9E37. This command does not affect the rate at
which programmed dc offset changes occur.

Command Syntax

[SOURce:]VOLTage:SLEW[:IMMediate]|INFinity

Parameters

1E-3 to 9.9E37 | INFinity |MINimum | MAXimum

Unit

V/S (volts per second)

*RST Value

INFinity

Examples

VOLT:SLEW 1 VOLT:SLEW MAX

VOLT:SLEW INF

Query Syntax

[SOURce:]VOLTage:SLEW[:IMMediate]?

Returned Parameters

Related Commands

VOLT:SLEW:MODE

VOLT:SLEW:TRIG