Appendix d: il series / hp6834b compatability – AMETEK Lx Series II Programming Manual User Manual
Page 235

Programming Manual - Rev S
Lx \ Ls Series II
231
Number
Message String
Cause
Remedy
814
"Voltage peak error "
Peak voltage
exceeds internal bus
voltage
This error may occur when selecting
user defined wave shapes with higher
crest factors. Reduce programmed
RMS value.
815
"Slew time exceed dwell" Time needed to slew
to final value is less
than dwell time.
Check dwell times in transient list
settings. Increase dwell time or
change slew rate for affected
parameter.
816
"Illegal during transient"
Operation requested
not available while
transient is running.
Wait till transient execution is
completed or abort transient
execution first.
817
"Output relay must be
closed"
Transient
programmed with
output relay open.
Close relay before attempting
transient operation.
819
"Clock and sync must be
internal"
Operation not
possible with external
clock
Switch to internal sync. (Default)
820
"Input buffer full"
Too much data
received.
Break up data in smaller blocks.
821
"Amplifier unbalance"
Hardware error. An
amplifier has an
overload condition.
Check amplifier balance adjustment.
If error persists contact CI service at
822
"Waveform harmonics
limit"
Harmonic contents of
user defined wave
shape is too high and
could damage
amplifier output
stage.
Reduce harmonic content or reduce
fundamental frequency programmed.
823
"Amplifier fault"
An amplifier failure.
Can be reported at
any time.
Determine which amplifier is at fault
with self-test or checking LED on
Relay Board. Replace amplifier.
824
“Auxiliary down”
One or more auxiliary
units is not powered
up or not working.
Turn on all auxiliary units.
825
“Over voltage prot trip”
Over voltage
detected on output
Check output voltage for correct RMS
value.
826
“Peak current prot trip”
Peak current limit
exceeded.
Peak current exceeded. Could be
caused by switching EUT on or off.
827
“Frequency error”
Frequency error
during self-test.
Correct frequency was not measured
during self-test. May be result of 801
error.
828
“Phase error”
Self test error phase
angle
Correct phase angle was not
measured during self-test. May be
result of 801 error.
829
“Dc component exceed
limit”
Too much DC
content in loaded
ARB waveform.
Check waveform programming.
Table 8-11: SCPI error codes and messages.
Appendix D: iL Series / HP6834B Compatability