Command Syntax

SYSTem:COMMunicate:LAN:DESCription

Parameters

Example

SYST:COMM:LAN:DESC “EVAL UNIT”

Query Syntax

SYST:COMM:LAN:DESC?

Returned Parameters

SYSTem:COMMunicate:LAN:GWADdress

This command will set the LAN Gateway address. The query format returns the Gateway address
set. To query the actual gateway address used by the LAN interface (in case the IP address is not
static but assigned by a DHCP server), the NCON parameter (Network Connection Setting) must
be added to the query. The default value is 0.0.0.0.

Command Syntax

SYSTem:COMMunicate:LAN:GWADddress

Parameters

ddd.ddd.ddd.ddd

Example

SYST:COMM:LAN:GWAD 255.255.255.0

Query Syntax

SYST:COMM:LAN:GWAD?

SYST:COMM:LAN:GWAD? NCON

Returned Parameters

SYSTem:COMMunicate:LAN:HBITs

This command will set the number of host address bits. The query format returns the number of
bits set. To query the actual number of bits used by the LAN interface (in case the IP address is
not static but assigned by a DHCP server), the NCON parameter (Network Connection Setting)
must be added to the query. The default value is 8.

Command Syntax

SYSTem:COMMunicate:LAN:HBITs

Parameters

8 to 32

Example

SYST:COMM:LAN:HBIT 8

Query Syntax

SYST:COMM:LAN: HBIT?

SYST:COMM:LAN: HBIT? NCON

Returned Parameters

SYSTem:COMMunicate:LAN:PASSword

This command will set the user password. The password is used to access to LAN configuration
screen from the Web page. The d

efault password is “admin”. The description is limited to 8

characters.

Command Syntax

SYSTem:COMMunicate:LAN:PASSword

Parameters

Example

SYST:COMM:LAN:PASS “admin”

Query Syntax

SYST:COMM:LAN:PASS?

Returned Parameters

SYSTem:COMMunicate:LAN:PORT

This command will set the socket port number. The query format returns the port number set. The
default value is 5025.

Command Syntax

SYSTem:COMMunicate:LAN:PORT

Parameters

0- 65535

Example

SYST:COMM:LAN:PORT 5025

Query Syntax

SYST:COMM:LAN: PORT?

Returned Parameters

SYSTem:COMMunicate:SERial:BAUD

This command will set the SERial communication baud rate.