4.4

Diagnostic Subsystem

These subsystem commands perform diagnostic functions which include reading and writing to
the EEPROM, resetting the AC source and reading temperature.

Subsystem Syntax

DIAGnostic

:RESet

Force power-on reset

:TEMPerature

:AMBient?

Returns ambient temperature in °C

DIAGnostic:RESet

This commands forces a power-on reset.

Command Syntax

DIAGnostic:RESet

Parameters

None

Examples

DIAG:RES

Related Commands

*RST

DIAGnostic:TEMPerature:AMBient?

This query returns the temperature measured at the ambient sense thermistor in degrees C.

Query Syntax

DIAGnostic:TEMPerature:AMBient?

Parameters

None

Examples

DIAG:TEMP:AMB?

Returned Parameters