Programming Manual

Lx \ Ls Series II

114

Command Syntax

ABORt

Parameters

None

Examples

ABOR

Related Commands

INIT

*RST

*TRG TRIG

INITiate:SEQuence
INITiate:NAME

The INITiate commands control the initiation of both the transient generator and the measurement
trigger systems. They cause the trigger system to make a transition from the Idle state to the
Waiting-for-Trigger state. If the trigger system is not in the Idle state, the initiate commands are
ignored.

INITiate:SEQuence and INITiate:NAME initiate the trigger systems to reference trigger
sequences. INITiate:SEQuence references a trigger sequence by its number, while
INITiate:NAME references a sequence by its name. The correspondence between sequence
names and numbers is:

Sequence Number

Sequence Name

Description

1

(default)

TRANsient

Step, pulse, or list transient trigger sequence

3

ACQuire

Measurement acquire trigger sequence

Command Syntax

INITiate[:IMMediate]:SEQuence[ 1 | 3 ]

INITiate[:IMMediate]:NAME

Parameters

For INIT:NAME TRANsient | ACQuire

Examples

INIT:SEQ1

INIT:NAME ACQ

Related Commands

ABOR INIT:CONT

TRIG

*TRG

INITiate:CONTinuous:SEQuence
INITiate:CONTinuous:NAME

These commands control the transient generator trigger system.

1 or ON continuously initiates the transient trigger system.

0 or OFF turns off continuous triggering. In this state, the trigger system must be initiated for
each triggered event using INITiate:SEQuence.

INITiate:CONTinuous:SEQuence references the transient trigger sequence by its number, while
INITiate:CONTinuous:NAME references it by its name.

Command Syntax

INITiate:CONTinuous:SEQuence[1]

INITiate:CONTinuous:NAME TRANsient,

Parameters

0|1|OFF|ON

*RST Value

OFF

Examples

INIT:CONT:SEQ ON

INIT:CONT:NAME TRAN, 1

Related Commands

ABOR INIT:CONT

TRIG

*TRG