Programming Manual

Lx \ Ls Series II

81

Command Syntax

[SOURce:]LIST:FREQuency[:LEVel]{,}

Parameters

45 to 5000

Unit

HZ (Hertz)

Examples

LIST:FREQ 60,65,70

Query Syntax

[SOURce:]LIST:FREQ?

Returned Parameters

Related Commands

LIST:FREQ:POIN? LIST:COUN LIST:DWEL LIST:STEP

LIST:FREQ:SLEW

LIST:FREQuency:POINts?

This query returns the number of points specified in LIST:FREQuency. Note that it returns only the
total number of points, not the point values.

Query Syntax

[SOURce:]LIST:FREQ[:LEVel]:POINts?

Returned Parameters

Example

LIST:FREQ:POIN?

Related Commands

LIST:FREQ

LIST:FREQuency:SLEW

This command sets the sequence of frequency slew list points. The frequency points are given in
the command parameters, which are separated by commas. The order in which the points are
entered determines the sequence in which they are output when a list is triggered. Changing list
data while a subsystem is in list mode generates an implied ABORt.

Command Syntax

[SOURce:]LIST:FREQuency:SLEW{,}

Parameters

0 to 9.9E31 | INFinity

Unit

HZ (Hertz) per second

Examples

LIST:FREQ:SLEW 10, 1E2, INF

Query Syntax

[SOURce:]LIST:FREQ:SLEW?

Returned Parameters

Related Commands

LIST:FREQ:SLEW:POIN? LIST:COUN LIST:DWEL LIST:STEP

LIST:FREQ

LIST:FREQuency:SLEW:POINts?

This query returns the number of points specified in LIST:FREQuency:SLEW. Note that it returns
only the total number of points, not the point values.

Query Syntax

[SOURce:]LIST:FREQ:SLEW:POINts?

Returned Parameters

Example

LIST:FREQ:SLEW:POIN?

Related Commands

LIST:FREQ:SLEW

LIST:MODE

This command sets the mode of operation for the transient list system. The default mode after
*RST is always BOT. In BOT mode, the output of the power source will return to the steady state
program values when the transient list execution completes. This mode is compatible with the
HP843B and CI iL series power sources. In EOT mode, the output of the power source will
remain at the final value reached at the end of the transient list execution. This mode is
compatible with other CI power source models. This command requires firmware revision 1.31 or
higher.

This mode is not stored in any setup register so the unit will always revert to BOT mode at power
up or after a *RST command.

Note that sending the ABORT command will always cause the output to revert to the last steady
state program values.