:FREQuency
[:LEVel]
:MODulation
:TRANsient
:VOLTage
[:LOW]
:HIGH
:FREQuency
[:LOW]
:HIGH
:VERSion D | E | F
8.6.2 MIL704 General Commands
MIL704:VERSion
This command will select the required test standard version. The version selection must be set
prior to any test selection.
Available revsions are D, E and F.
Command Syntax
MIL704:VERSion
Parameters
D | E | F
Examples
MIL704:VERS E
Query Syntax
MIL704:VERS?
Returned Parameters
MIL704:FREQuency
This command will selects the nominal test frequency. For revisions D and E, the only available
setting is 400 Hz. For revision F (draft), available settings are 60 Hz, 400 Hz or variable frequency.
The nominal frequency selection must be set prior to any test execution.
Command Syntax
MIL704:FREQuency
Parameters
60HZ | 400HZ | VFRequency
Examples
MIL704:FREQ VFR
Query Syntax
MIL704:FREQ?
Returned Parameters