Lx \ Ls Series II

181

SRQ 1

SRQ 1

Reported Message

Cause

90

26

RNG RANGE ERROR

RNG value greater than highest range

90

27

AMP RANGE ERROR

AMP value greater than RNG value

90

28

FRQ RANGE ERROR

FRQ value is less than 45 or greater than 1000
Hz (Ls) or 5000 Hz (Ls-HF)

93

29

PHZ RANGE ERROR

PHZ value greater than ±999.0

94

30

CRL RANGE ERROR

CRL value greater than maximum value

95

31

RMPA RANGE ERROR

DLY, STP or VAL values wrong

96

32

SYNTAX ERROR

Wrong string syntax

97

33

BUS LOCAL ERROR

Remote message sent while in local mode

98

34

EXT SYNC ERROR

No external sync input or signal not between
min and max frequency

99

35

CPU MEMORY FAULT

CPU failed power-on self-test

100

36

DMA OVERFLOW ERROR

Remote message greater than 256 bytes

63

63

The response after SRQ2 is included in a setup
string and the execution of the string or
measurement is completed.

40

40

STA OK

No Error

Table 8-6: APE Status Byte Error Codes