Programming Manual

Lx \ Ls Series II

108

Command Syntax

SYSTem:COMMunicate:SERial:BAUD

Parameters

9600 | 19200 | 38400 | 57600 | 115200 | 230400 | 460800

Example

SYST:COMM:SER:BAUD 38400

Returned Parameters

SYSTem:COMMunicate:SERial:PARity

This command will set the SERial communication parity.

Command Syntax

SYSTem:COMMunicate:SERial:PARity

Parameters

NONE|EVEN|ODD

Example

SYST:COMM:SER:PAR NONE

Returned Parameters

SYSTem:COMMunicate:SERial:BITS

This command will set the SERial communication number of data bits.

Command Syntax

SYSTem:COMMunicate:SERial:BITS

Parameters

7|8

Example

SYST:COMM:SER:BITS 8

Returned Parameters

SYSTem:COMMunicate:SERial:SBITs

This command will set the SERial communication number of stop bits.

Command Syntax

SYSTem:COMMunicate:SERial:SBITs

Parameters

1|2

Example

SYST:COMM:SER:SBITs 1

Returned Parameters

SYSTem:CONFiguration

This query format of this command returns the installed options and configuration settings of the
power source. It is not possible to change the configuration as only the query form is supported.

Command Syntax

SYSTem:CONFigure?

Query Syntax

SYSTem:CONFigure?

Returned Parameters

Related Commands

*OPT?

SYSTem:CONFiguration:NOUTputs

This command selects the number of outputs phases for the AC source. This requires that the AC
source is capable of switching between single and three-phase mode.

Note: Execution of this command disables all outputs, clears lists and *RCL states to the

initialization default values, reconfigures current readback and programming calibration
constants.

Once configured, the instrument behaves transparently as either a single-phase source or as a
three phase source depending on the selected configuration.