AMETEK Lx Series II Programming Manual User Manual
Page 112

Programming Manual
Lx \ Ls Series II
108
Command Syntax
SYSTem:COMMunicate:SERial:BAUD
Parameters
9600 | 19200 | 38400 | 57600 | 115200 | 230400 | 460800
Example
SYST:COMM:SER:BAUD 38400
Returned Parameters
SYSTem:COMMunicate:SERial:PARity
This command will set the SERial communication parity.
Command Syntax
SYSTem:COMMunicate:SERial:PARity
Parameters
NONE|EVEN|ODD
Example
SYST:COMM:SER:PAR NONE
Returned Parameters
SYSTem:COMMunicate:SERial:BITS
This command will set the SERial communication number of data bits.
Command Syntax
SYSTem:COMMunicate:SERial:BITS
Parameters
7|8
Example
SYST:COMM:SER:BITS 8
Returned Parameters
SYSTem:COMMunicate:SERial:SBITs
This command will set the SERial communication number of stop bits.
Command Syntax
SYSTem:COMMunicate:SERial:SBITs
Parameters
1|2
Example
SYST:COMM:SER:SBITs 1
Returned Parameters
SYSTem:CONFiguration
This query format of this command returns the installed options and configuration settings of the
power source. It is not possible to change the configuration as only the query form is supported.
Command Syntax
SYSTem:CONFigure?
Query Syntax
SYSTem:CONFigure?
Returned Parameters
Related Commands
*OPT?
SYSTem:CONFiguration:NOUTputs
This command selects the number of outputs phases for the AC source. This requires that the AC
source is capable of switching between single and three-phase mode.
Note: Execution of this command disables all outputs, clears lists and *RCL states to the
initialization default values, reconfigures current readback and programming calibration
constants.
Once configured, the instrument behaves transparently as either a single-phase source or as a
three phase source depending on the selected configuration.