OUTPut:PROTection:DELay

This command sets the delay time between the programming of an output change that produces
a CL or UNREG status condition and the recording of that condition by the Status Operation
Condition register. The delay prevents momentary changes in status that can occur during
programming from being registered as events by the status subsystem. In most cases these
temporary conditions are not considered an event, and to record them as such would be a
nuisance.

Command Syntax

OUTPut:PROTection:DELay

Parameters

0 to 32 | MINimum | MAXimum

Unit

S (seconds)

*RST Value

100 milliseconds

Examples

OUTP:PROT:DEL 75E-1

Query Syntax

OUTPut:PROTection:DELay?

Returned Parameters

Related Commands

OUTP:PROT:CLE

*RCL

*SAV

OUTPut:RI[:LEVel]

This command sets the remote inhibit level mode. Factory default is LOW, which requires a
contact closure to open the output relay. The level can be reversed by setting it to HIGH. Once
set, the RI level setting is retained each time the power source is powered up. Note that this
command is only implement with firmware revision 1.69 or higher. Lower firmware revisions only
provide the default LOW setting.

Command Syntax

OUTPut:RI:LEVel

Parameters

LOW | HIGH

*RST Value

LOW

Examples

OUTP:RI:LEV HIGH

Query Syntax

OUTP:RI:LEV?

Returned Parameters

Related Commands

OUTP OUTP:RI:MODE

OUTPut:RI:MODE

This command selects the mode of operation of the Remote Inhibit protection. The following
modes can be selected:

LATChing

A TTL low at the RI input latches the output in the protection shutdown
state, which can only be cleared by OUTPut:PROTection:CLEar.

LIVE

The output state follows the state of the RI input. A TTL low at the RI
input turns the output off; a TTL high turns the output on.

OFF

The instrument ignores the RI input.

The RI output state is saved as part of an instrument setup using the *SAV command. It can be
made part of the power on setting if needed. The default state is LIVE.

Command Syntax

OUTPut:RI:MODE

Parameters

LATChing | LIVE | OFF

*RST Value

OFF

Examples

OUTP:RI:MODE LIVE

Query Syntax

OUTPut:RI:MODE?

Returned Parameters

Related Commands

OUTP:PROT:CLE

OUTP:RI:LEV