Programming Manual

Lx \ Ls Series II

50

Query Syntax

MEASure:ARRay:VOLTage[:DC]? [, ]

FETCh:ARRay:VOLTage[:DC]? [, ]

Parameters

Optional block and offset parameters ,. Where the first value
is the number of 256 sample blocks to transfer and the second value
is the first block (offset) to start with. Number of blocks is from 1 to 16,
offset is from 0 to 15.

Examples

MEAS:ARR:VOLT?

FETC:ARR:VOLT?

Returned Parameters

4096 or less NR3 values

Related Commands

INST:NSEL

SENS:SWE

MEASure:ARRay:VOLTage:HARMonic?
FETCh:ARRay:VOLTage:HARMonic?

Phase Selectable

These queries return an array of harmonic amplitudes of output voltage in rms volts. The first
value returned is the dc component, the second value is the fundamental frequency, and so on up
to the 50th harmonic. Harmonic orders can be measured up to the fundamental measurement
bandwidth of the measurement system, which is 16 kHz. Thus, the maximum harmonic that can
be measured is dependent on the output frequency. Any harmonics that represent frequencies
greater than 16 kHz are returned as 0.

Query Syntax

MEASure:ARRay:VOLTage:HARMonic[:AMPLitude]?

FETCh:ARRay:VOLTage:HARMonic[:AMPLitude]?

Parameters

None

Examples

MEAS:ARR:VOLT:HARM?

FETC:ARR:VOLT:HARM?

Returned Parameters

51 NR3 values

Related Commands

INST:NSEL