AMETEK Lx Series II Programming Manual User Manual
Page 232
![background image](https://www.manualsdir.com/files/771680/content/doc232.png)
Programming Manual - Rev S
Lx \ Ls Series II
228
Number
Message String
Cause
Remedy
-254
"Media full"
No storage space left
to save settings or
data.
Delete other settings or data to make
room.
-255
“Directory full”
Too many waveform
directory entries
Delete one or more waveforms from
waveform memory to make room.
-256
“File name not found”
Waveform requested
not in directory
Check waveform directory for
waveform names present.
-257
“File name error”
Incorrect filename
Too many or non ASCII characters
used in waveform file definition.
-283
“Illegal variable name”
Variable name illegal.
Use ASCII characters only
-300
"Device specific error"
Hardware related
error
Check hardware for proper operation.
-311
"Memory error"
Waveform memory
checksum error.
May be the result of incomplete user-
defined waveform download. Check
interface and try downloading
waveform again. Successful
download may clear this error
condition.
Alternatively, use TRAC:DEL:ALL
command to clear waveform memory.
-314
"Save/recall memory
lost"
User setup register
contents lost
Store setup in same register again.
-315
"Configuration memory
lost"
Hardware
configuration settings
lost.
Contact AMETEK service department
at
to obtain instructions on restoring
configuration data.
-330
"Self-test failed"
Internal error
Contact AMETEK service department
at
-350
"Queue overflow"
Message queue full.
Too many message. Read status
using SYST:ERR query until 0, "No
Error" is received indicating queue
empty.
-400
"Query error"
Unable to complete
query.
Check programming manual for
correct query format and parameters
-410
"Query INTERRUPTED"
Query issued but
response not read.
Check application program for correct
flow. Response must be read after
each query to avoid this error.
-420
"Query
UNTERMINATED"
Query incomplete.
Check for terminator after query
command.
-430
"Query DEADLOCKED"
Query cannot be
completed
Check application program for
multiple queries
-440
"Query
UNTERMINATED"
Query incomplete.
Check for terminator after query
command.
0
"No error"
No errors in queue
2
" Non-volatile RAM
CONFIG section
checksum failed"
Controller failure
during Self-test.
Contact AMETEK service department
at
3
" Non-volatile RAM CAL
section checksum failed"
Controller failure
during Self-test.
Contact AMETEK service department
at
4
" Non-volatile RAM
WAVEFORM section
Controller failure
during Self-test.
Contact AMETEK service department
at