Programming Manual - Rev S
Lx \ Ls Series II
230
Number
Message String
Cause
Remedy
604
"Trigger received before
requested number of
pre-trigger readings"
Data acquisition pre-
trigger buffer not
filled yet.
Hold off trigger or reduce pre-trigger
delay.
605
"Requested RMS current
too high for voltage
range"
Max RMS current is
function of voltage
range selected.
Reduce programmed RMS current
limit or select low voltage range.
606
"Waveform data not
defined"
No waveform name
specified
Specify waveform name before
sending waveform data.
607
"VOLT,VOLT:SLEW, and
FUNC:SHAPe modes
incompatible"
Conflict between
wave shape and
programmed slew
Reduce slew or change waveform
type.
608
"Measurement
overrange"
Measurement data
out of range.
609
"Output buffer overrun"
Too much data in
output buffer.
Check receive mode on application
program. Program is not reading data
sent by AC source.
610
"Command cannot be
given with present
SYST:CONF setting"
Command conflicts
with available
hardware or firmware
option settings.
Check configuration for available
options and features.
801
"Output volt fault"
- Output voltage does
not match program
value when ALC is
on.
- Over load
- Voltage kick-back
- No output voltage
Load exceeds current limit and unit is
in Constant Voltage (CV) mode of
operation. Reduce load or increase
CL setting.
Output voltage is driven above
programmed voltage by external
influence (Load, voltage kickback,
etc.)
802
"Current limit fault"
Current limit
exceeded.
Load exceeds current limit and unit is
in Constant Voltage (CV) mode of
operation. Reduce load or increase
CL setting
803
"Temperature fault"
Amplifier heat sink
temp. too high.
Reduce load. Ensure proper air flow
and exhaust clearance. Check fan(s)
for operation.
804
"External sync. error"
Could not sync to
external sync signal.
External sync signal missing,
disconnected or out of range.
805
"Initial memory lost"
Initial settings could
not be recalled at
power-up.
Save power on settings again to
overwrite old content.
806
"Limit memory lost"
Hardware
configuration settings
could not be recalled
at power-up.
Contact AMETEK service department
at
to obtain instructions on restoring
configuration data.
807
"System memory lost"
Memory corrupted
during power-up.
Recycle power.
808
"Calibration memory lost" Calibration data lost
during power-up.
Contact AMETEK service department
at
to obtain instructions on restoring
calibration data or recalibrate unit.
813
"Missing list parameter"
One or more
transient list
parameters missing.
Check programmed lists.