Programming Manual

Lx \ Ls Series II

38

Command Syntax

CALibrate:MEASure:VOLTage

Parameters

Reference instrument voltage reading

Query Syntax

CALibrate:MEASure:VOLTage?

Returned Parameters

Examples

CAL:MEAS:VOLT

Related Commands

CAL:SAVE

CAL:MEAS:CURR

CALibrate:PASSword

This command can only be used to unlock the calibration mode. Once unlocked, non-query
calibration commands will be accepted. Query commands are always accepted.

Command Syntax

CALibrate:PASSword

Parameters

(default)

Examples

CAL:PASS 300

Related Commands

none

CALibrate:SAVE

This command can only be used in calibration mode. It saves any new calibration constants (after
a current or voltage calibration procedure has been completed) in nonvolatile memory.

Command Syntax:

CALibrate:SAVE

Parameters

None

Examples

CAL:SAVE

Related Commands

CAL:CURR

CAL:VOLT

CALibrate[:SOURce]:PHASe

Phase Selectable

This command can be used to set the phase calibration offset coefficient. Use the INST:SEL or
INST:NSEL to select the desired phase. This allows the phase for voltage 2 and 3 (B and C) to
be adjusted with respect to phase A. The query format returns the actual calibration coefficient

Command Syntax

CALibrate[:SOURce]:PHASe

Parameters

Query Syntax

CALibrate[:SOURce]:PHASe?

Returned Parameters

Examples

CAL:PHAS 1.3

Related Commands

none

CALibrate[:SOURce]:VOLTage:HRANge

Phase Selectable

This command can be used to query the output voltage calibration coefficient. Only the query
format is available. Use the INST:SEL or INST:NSEL to select the desired phase. The parameter
can be any integer value between 0 and 4000. Factory default is 450.

Command Syntax

CALibrate[:SOURce]:VOLTage:HRANge

Parameters

0 to 4000

Query Syntax

CALibrate[:SOURce]:VOLTage:HRANge?

Returned Parameters

Examples

CAL:VOLT:HRAN 450

Related Commands

none