Lx \ Ls Series II

208

8.7.3 MS704 Steady State Commands

Steady state tests can be started by sending one of the MS704:NORM commands. Available
normal state commands are listed below.

MS704:NORMal:SSLimits[:ALL]

This command will run all steady state normal operation mode tests.

Command Syntax

MS704:NORMal:SSLimits[:ALL]

Parameters

none

Examples

MS704:NORM:SSL

Query Syntax

none

MS704:NORMal:SSLimits:NOMinal

This command will run the steady state normal operation nominal voltage test.

Command Syntax

MS704:NORMal:SSLimits:NOMinal

Parameters

none

Examples

MS704:NORM:SSL:NOM

Query Syntax

none

MS704:NORMal:SSLimits:LOW

This command will run the steady state normal operation low voltage test.

Command Syntax

MS704:NORMal:SSLimits:LOW

Parameters

none

Examples

MS704:NORM:SSL:LOW

Query Syntax

none

MS704:NORMal:SSLimits:HIGH

This command will run the steady state normal operation high voltage test.

Command Syntax

MS704:NORMal:SSLimits:HIGH

Parameters

none

Examples

MS704:NORM:SSL:HIGH

Query Syntax

none

MS704:NORMal:SSLimits:UNBalance

This command will run the steady state normal operation voltage unbalance test. This test is only
available in the 3 phase mode.

Command Syntax

MS704:NORMal:SSLimits:UNBalance

Parameters

none

Examples

MS704:NORM:SSL:UNB

Query Syntax

none