AMETEK Lx Series II Programming Manual User Manual

Page 198

background image

Programming Manual

Lx \ Ls Series II

194

DO160:ABNormal:FREQuency:TRANsient

This command applies to Group 1 only for the EuroCAE standard. Refer to the User Manual for
details.

Command Syntax

DO160:ABNormal:FREQuency:TRANsient

Parameters

none

Examples

DO160:ABN:FREQ:TRAN

Query Syntax

none

Returned Parameters

none

8.4.5 DO160 Test Setup Commands

DO160:STANdard

This command will select the proper test standard. The standard selection must be set prior to
any test selection.

The RTCA2 selection includes the Change 2 notice to the DO160D test standard as it applies to
section 16. The RTCA selection does not include the Change 2 updates.

The Airbus standard is the same as the EuroCAE-14D standard with the exception of nominal
voltage modulation.

Command Syntax

DO160:STANdard

Parameters

RTCA | RTCA2 | EURocae | AIRBus

Examples

DO160:STAN RTCA

Query Syntax

DO160:STAN?

DO160:GROup

This command is used to select equipment Group 1, 2 or 3 for the EUROCAE standard or
equipment Group ACF, ANF or AWF for the RTCA2 standard. Some test (commands) apply only
to specific groups of equipment. If a test is executed for a group to which it does not apply, the
command is ignored by the power source. Note that in this case no error is generated. The
application program is responsible for selecting the relevant Standard, Group and test.

Command Syntax

DO160:GROup

Parameters

1 | 2 | 3 | ACF | ANF | AWF

Examples

DO160:GROup 2

Query Syntax

DO160:GRO?