AMETEK Lx Series II Programming Manual User Manual
Page 198

Programming Manual
Lx \ Ls Series II
194
DO160:ABNormal:FREQuency:TRANsient
This command applies to Group 1 only for the EuroCAE standard. Refer to the User Manual for
details.
Command Syntax
DO160:ABNormal:FREQuency:TRANsient
Parameters
none
Examples
DO160:ABN:FREQ:TRAN
Query Syntax
none
Returned Parameters
none
8.4.5 DO160 Test Setup Commands
DO160:STANdard
This command will select the proper test standard. The standard selection must be set prior to
any test selection.
The RTCA2 selection includes the Change 2 notice to the DO160D test standard as it applies to
section 16. The RTCA selection does not include the Change 2 updates.
The Airbus standard is the same as the EuroCAE-14D standard with the exception of nominal
voltage modulation.
Command Syntax
DO160:STANdard
Parameters
RTCA | RTCA2 | EURocae | AIRBus
Examples
DO160:STAN RTCA
Query Syntax
DO160:STAN?
DO160:GROup
This command is used to select equipment Group 1, 2 or 3 for the EUROCAE standard or
equipment Group ACF, ANF or AWF for the RTCA2 standard. Some test (commands) apply only
to specific groups of equipment. If a test is executed for a group to which it does not apply, the
command is ignored by the power source. Note that in this case no error is generated. The
application program is responsible for selecting the relevant Standard, Group and test.
Command Syntax
DO160:GROup
Parameters
1 | 2 | 3 | ACF | ANF | AWF
Examples
DO160:GROup 2
Query Syntax
DO160:GRO?