Programming Manual

Lx \ Ls Series II

63

OUTPut:DFI:SOURce

This command selects the source for DFI events. The choices are:

QUEStionable

Questionable summary bit

OPERation

Operation summary bit

ESB

Standard Event summary bit

RQS

Request Service summary bit

OFF

Never true

Command Syntax

OUTP:DFI:SOUR

Parameters

QUES | OPER | ESP | RQS | OFF

*RST Value

OFF

Examples

OUTP:DFI:SOUR OPER

Query Syntax

OUTPut:DFI:SOUR?

Returned Parameters

Related Commands

OUTP:DFI

OUTPut:PON[:STATe]

This command selects the power-on state of the AC source. The following states can be selected:

RST

Sets the power-on state to *RST. Refer to the *RST command as
described later in this chapter for more information.

RCL0

Sets the power-on state to *RCL 0. Refer to the *RCL command as
described later in this chapter for more information.

Command Syntax

OUTPut:PON[:STATE]

Parameters

RST | RCL0

Examples

OUTP:PON:STAT RST

Query Syntax

OUTPut:PON:STATe?

Returned Parameters

Related Commands

*RST

*RCL

OUTPut:PROTection:CLEar

This command clears the latch that disables the output when an overvoltage (OV), overcurrent
(OC), overtemperature (OT), or remote inhibit (RI) fault condition is detected. All conditions that
generated the fault must be removed before the latch can be cleared. The output is then restored
to the state it was in before the fault condition occurred.

Command Syntax

OUTPut:PROTection:CLEar

Parameters

None

Examples

OUTP:PROT:CLE

Related Commands

OUTP:PROT:DEL

*RCL

*SAV